Effect of deformable registration uncertainty on lung SBRT dose accumulation
- 23 December 2015
- journal article
- Published by Wiley in Medical Physics
- Vol. 43 (1) , 233-240
Abstract
Purpose: Deformable image registration (DIR) plays an important role in dose accumulation, such as incorporating breathing motion into the accumulation of the delivered dose based on daily 4DCBCT images. However, it is not yet well understood how the uncertainties associated with DIR methods affect the dose calculations and resulting clinical metrics. The purpose of this study is to evaluate the impact of DIR uncertainty on the clinical metrics derived from its use in dose accumulation. Methods: A biomechanical model based DIR method and a biomechanical-intensity-based hybrid method, which reduced the average registration error by 1.6 mm, were applied to ten lungcancer patients. A clinically relevant dose parameter [minimum dose to 0.5 cm3 (Dmin)] was calculated for three dose scenarios using both algorithms. Dose scenarios included static (no breathing motion), predicted (breathing motion at the time of planning), and total accumulated (interfraction breathing motion). The relationship between the dose parameter and a combination of DIR uncertainty metrics, tumor volume, and dose heterogeneity of the plan was investigated. Results: Depending on the dose heterogeneity, tumor volume, and DIR uncertainty, in over 50% of the patients, differences greater than 1.0 Gy were observed in the Dmin of the tumor in the static dose calculation on exhale phase of the 4DCT. Such differences were due to the errors in propagating the tumor contours from the reference planning 4DCT phase onto a subsequent 4DCT phase using each DIR algorithm and calculating the dose on that phase. The differences in predicted dose were more subtle when breathing motion was modeled explicitly at the time of planning with only one patient exhibiting a greater than 1.0 Gy difference in Dmin. Dmin differences of up to 2.5 Gy were found in the total accumulated delivered dose due to difference in quantifying the interfraction variations. Such dose uncertainties could potentially be clinically significant. Conclusions: Reductions in average uncertainty in DIR algorithms by 1.6 mm may have a clinically significant impact on the decision-making metrics used in dose planning and dose accumulation assessment.Keywords
Funding Information
- National Institutes of Health (NIH) (5RO1CA124714)
- National Institutes of Health (5RO1CA124714)
- National Institutes of Health (5RO1CA124714)
This publication has 23 references indexed in Scilit:
- Voxel-based statistical analysis of uncertainties associated with deformable image registrationPhysics in Medicine & Biology, 2013
- On the dosimetric effect and reduction of inverse consistency and transitivity errors in deformable image registration for dose accumulationMedical Physics, 2011
- Accurate Accumulation of Dose for Improved Understanding of Radiation Effects in Normal TissueInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Cumulative Lung Dose for Several Motion Management Strategies as a Function of Pretreatment Patient ParametersInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Dense image registration through MRFs and efficient linear programmingMedical Image Analysis, 2008
- Contact surface and material nonlinearity modeling of human lungsPhysics in Medicine & Biology, 2007
- Deformable registration of the planning image (kVCT) and the daily images (MVCT) for adaptive radiation therapyPhysics in Medicine & Biology, 2006
- Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and Monte Carlo simulationsPhysics in Medicine & Biology, 2006
- Dose reconstruction in deforming lung anatomy: Dose grid size effects and clinical implicationsMedical Physics, 2005
- Inclusion of organ deformation in dose calculationsMedical Physics, 2003