Varying Effects of Corneal Thickness on Intraocular Pressure Measurements with Different Tonometers
Authors:
Open Access
- 16 July 2004
- journal article
- research article
- Published by Springer Nature in Eye
- Vol. 19 (3) , 327-332
Abstract
To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings. Ultrasound pachymetry and tonometry were performed on 170 eyes. Using the corrected GAT values that took CCT into account as the standard, we calculated the NCT and OBFT measurement errors related to the CCT variable by a linear regression model. Group comparisons were performed with the χ2 test and one-way ANOVA test. The correlation between the various tonometer measurements was analysed by Pearson's correlation method. Both the NCT (r=0.872, Pr=0.861, PPβ=0.063, r=0.650) and the GAT measurements the least (β=0.037, r=0.496). A linear regression model indicated that a 10 μm change in CCT resulted in a NCT measurement deviation of 0.47–0.98 mmHg and an OBFT measurement deviation of 0.29–0.81 mmHg. Pressure readings with the GAT, NCT, and OBFT are all affected by CCT, with the NCT being the one most affected and the GAT the least. Our findings suggest CCT an essential variable to consider in interpreting IOP readings, especially for the NCT measurements.Keywords
This publication has 21 references indexed in Scilit:
- Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non‐contact techniques: Reykjavik Eye StudyActa Ophthalmologica Scandinavica, 2002
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- Reliability of intraocular pressure measurements after myopic excimer photorefractive keratectomy11The authors have no proprietary interest in these or any other tonometers.Ophthalmology, 1998
- Ocular Hypertension and Central Corneal ThicknessOphthalmology, 1995
- Validity of pulsatile ocular blood flow measurementsSurvey of Ophthalmology, 1994
- The Effect of Corneal Thickness on Applanation TonometryAmerican Journal of Ophthalmology, 1993
- Increased Corneal Thickness Simulating Elevated Intraocular PressureArchives of Ophthalmology (1950), 1978
- APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESSActa Ophthalmologica, 1975
- A Noncontact Applanation TonometerArchives of Ophthalmology (1950), 1974
- Über ApplanationstonometrieOphthalmologica, 1957