Experimental validation of a method characterizing bow tie filters in CT scanners using a real‐time dose probe
- 18 February 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 38 (3) , 1406-1415
Abstract
Purpose: Beam‐shaping or “bow tie” (BT) filters are used to spatially modulate the x‐ray beam in a CT scanner, but the conventional method of step‐and‐shoot measurement to characterize a beam's profile is tedious and time‐consuming. The theory for characterization of bow tie relative attenuation (COBRA) method, which relies on a real‐time dosimeter to address the issues of conventional measurement techniques, was previously demonstrated using computer simulations. In this study, the feasibility of the COBRA theory is further validated experimentally through the employment of a prototype real‐time radiation meter and a known BT filter.Methods: The COBRA method consisted of four basic steps: (1) The probe was placed at the edge of a scanner's field of view; (2) a real‐time signal train was collected as the scanner's gantry rotated with the x‐ray beam on; (3) the signal train, without a BT filter, was modeled using peak values measured in the signal train of step 2; and (4) the relative attenuation of the BT filter was estimated from filtered and unfiltered data sets. The prototype probe was first verified to have an isotropic and linear response to incident x‐rays. The COBRA method was then tested on a dedicated breast CT scanner with a custom‐designed BT filter and compared to the conventional step‐and‐shoot characterization of the BT filter. Using basis decomposition of dual energy signal data, the thickness of the filter was estimated and compared to the BT filter's manufacturing specifications. The COBRA method was also demonstrated with a clinical whole body CT scanner using the body BT filter. The relative attenuation was calculated at four discrete x‐ray tube potentials and used to estimate the thickness of the BT filter.Results: The prototype probe was found to have a linear and isotropic response to x‐rays. The relative attenuation produced from the COBRA method fell within the error of the relative attenuation measured with the step‐and‐shoot method. The BT filter thickness estimates resulting from the dual energy scans on the breast CT system were equivalent to the manufacturing specifications. The clinical CT evaluation produced data conceptually similar to previous computer simulations and plausible relative attenuation profiles were observed.Conclusions: The COBRA method is a fast and accurate method for BT filter characterization, which requires a simple experimental setup in a clinical environment. Because of the ease of data acquisition, multienergy scans can be acquired which allow characterization of the BT filter thickness.Keywords
Funding Information
- National Institutes of Health (R01 EB002138)
This publication has 17 references indexed in Scilit:
- Method for evaluating bow tie filter angle‐dependent attenuation in CT: Theory and simulation resultsMedical Physics, 2009
- The development, validation and application of a multi-detector CT (MDCT) scanner model for assessing organ doses to the pregnant patient and the fetus using Monte Carlo simulationsPhysics in Medicine & Biology, 2009
- The influence of bowtie filtration on cone‐beam CT image qualityMedical Physics, 2008
- Compensators for dose and scatter management in cone‐beam computed tomographyMedical Physics, 2007
- Development and validation of MCNP4C-based Monte Carlo simulator for fan- and cone-beam x-ray CTPhysics in Medicine & Biology, 2005
- A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantomsPhysics in Medicine & Biology, 2005
- Simulation of CT dose and contrast-to-noise as function of bowtie shape2004
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- Generalized image combinations in dual KVP digital radiographyMedical Physics, 1981
- Energy-selective reconstructions in X-ray computerised tomographyPhysics in Medicine & Biology, 1976