CO2 Lasers and Temperature Changes of Titanium Implants
Authors:
- 1 December 1995
- journal article
- research article
- Published by Wiley in Journal of Periodontology
- Vol. 66 (12) , 1017-1024
Abstract
Lasers may be useful in uncovering submerged implants or in removing contaminants from “ailing” implants. The purposes of this study were to record temperature changes at the bone‐titanium implant interface when using a CO2 laser to: 1) uncover the implant at second stage surgery; and 2) “decontaminate” exposed implant threads. Scanning electron microscopy (SEM) was used to characterize surface changes of lased implants, both uncontaminated or contaminated with blood or saliva. In part one, 28 titanium implants, measuring 3.75 mm by 7 to 20 mm, were placed into room temperature, fresh, resected pig mandibles and covered with a flap of gingiva. The overlying tissue was removed with a CO2 laser at different power levels. Bone‐implant interface temperatures were measured with a thermocouple near the top of the implant, and 5 to 7 mm apical to the osseous crest. The effects of implant size, power level, tissue thickness, and operation time were evaluated. In part two, 5 mm by 4 mm bony dehiscences were created on 3 implants in one mandible and the exposed fixture threads lased at varying times and power levels. The results from part one showed temperature increases at the top thermocouple ranged from 4.2 to 16.8°C and increases at the bottom thermocouple ranged from 2.0 to 11.5°C. The results from part two showed temperature increases at the top thermocouple ranged from 1.2 to 11.7°C and increases at the bottom thermocouple from 0.0 to 5.0°C. If baseline ambient temperatures are 37°C, then the temperature at the bone‐implant interface might exceed 50°C. SEM revealed no gross surface changes in lased uncontaminated implants, but laser treatment alone of contaminated implants failed to completely remove saliva or blood. Further study is needed regarding temperature increases and surface changes induced by lasers that may adversely affect osseointegration. J Periodontol 1995;66:1017–1024.Keywords
This publication has 25 references indexed in Scilit:
- Current Status of Lasers in Soft Tissue Dental SurgeryJournal of Periodontology, 1993
- Bone ablation with Er:YAG and CO2 laser: Study of thermal and acoustic effectsLasers in Surgery and Medicine, 1992
- Comparison of three lasers for dental instrument sterilizationLasers in Surgery and Medicine, 1991
- CO2 laser application to the mineralized dental tissues—the possibility of iatrogenic sequelaeJournal of Dentistry, 1990
- The heat generated on the external root surface during post space preparationInternational Endodontic Journal, 1989
- Laser reflectance: Hazard in the dental operatoryOral Surgery, Oral Medicine, Oral Pathology, 1988
- Use of the CO2 laser in soft tissue dental surgeryLasers in Surgery and Medicine, 1987
- Latest treatment in dentistry by means of the CO2 laser beamLasers in Surgery and Medicine, 1986
- A new method of sterilization: the carbon dioxide laserJournal of Oral Pathology & Medicine, 1979