Technical Note: Long‐term monitoring of diode sensitivity degradation induced by proton irradiation
- 5 October 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (11) , 6634-6641
Abstract
Purpose: To measure diode sensitivity degradation (DSD) induced by cumulative proton dose delivered to a commercial daily quality assurance (QA) device.Methods: At our institution, six Daily QA 3 (DQA3, Sun Nuclear Corporation, Melbourne, FL, USA) devices have been used for daily proton pencil beam scanning QA in four proton gantry rooms over a span of 4 years. DQA3 diode counts were cross‐calibrated using a homogenous field with a known dose of 1 Gy. The DSD rate (ΔR%/100 Gy) was calculated using linear regression on time‐series plots of diode counts and an estimate of cumulative dose per year based on the cross‐calibration. The effect of DSD on daily QA spot position measurements was quantified by converting DSD to baseline spot position shift.Results: The average dose delivered to the four inner DQA3 diodes was 104 ± 5 Gy/year, and the rate of DSD was ‐5.1% ± 1.0/100 Gy with the exception of one DQA3 device that had a significantly higher rate of DSD (–12%/100 Gy). The R2s of the linear fit to time‐series plots were between 0.92 and 0.98. The DSD rates were not constant but decreases with accumulated doses. The four center diodes, which received 40% of the cumulative dose received by inner diodes, had a DSD rate of –7.2% ± 0.9/100 Gy. For our daily QA program, 1 year of DSD was equivalent to a 0.2 mm shift in spot position.Conclusions: The DSD rate of DQA3 diodes determined by long‐term proton daily QA data was about –5%/100 Gy, which is more than 10 times greater than the reported DSD rate from photon irradiation. DQA3 diodes may be used for daily proton QA programs, provided that they are recalibrated at an appropriate frequency that should be determined specifically for different daily QA programs.Keywords
This publication has 15 references indexed in Scilit:
- Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton BeamsCancers, 2015
- Daily QA in proton therapy using a single commercially available detectorJournal of Applied Clinical Medical Physics, 2014
- Patient‐specific QA and delivery verification of scanned ion beam at NIRS‐HIMACMedical Physics, 2013
- A novel daily QA system for proton therapyJournal of Applied Clinical Medical Physics, 2013
- Patient-Specific Quality Assurance for Prostate Cancer Patients Receiving Spot Scanning Proton Therapy Using Single-Field Uniform DoseInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center–HoustonMedical Physics, 2009
- Evaluation of a 2D diode array for IMRT quality assuranceRadiotherapy and Oncology, 2004
- A 2-D diode array and analysis software for verification of intensity modulated radiation therapy deliveryMedical Physics, 2003
- Characteristics of silicon and diamond detectors in a 60 MeV proton beamPhysics in Medicine & Biology, 2002