Cryo-Planing Biological Specimens for Scanning Electron Microscopy
- 14 March 2019
- journal article
- research article
- Published by Oxford University Press (OUP) in Microscopy Today
- Vol. 27 (2) , 20-25
Abstract
Cryo-planing SEM provides a powerful 3D view of internal and external tissue and cell features. However, specimen preparation is tedious because it requires custom apparatus and difficult cryogenic manipulations. We present an easy method using mPrep/s capsules that provide a “handle” to hold and orient specimens throughout cryo-preparation and imaging. Cryo-facing is done with a cryo-ultramicrotome, requiring no custom equipment while providing accurate control of the image plane. We show this with fresh Christmas cactus leaves and aldehyde-fixed kidney tissue imaged by room temperature SEM after freeze-drying, demonstrating the method operating without an expensive cryo-SEM.Keywords
This publication has 7 references indexed in Scilit:
- Cryo-Planing Vs Freeze Fracture: Sample Preparation for Cryo-HRSEMMicroscopy and Microanalysis, 2015
- Capsule-Based Processing and Handling of Electron Microscopy Specimens and GridsMicroscopy Today, 2015
- Cryo-Planing of Small Biological Specimens for SEM Using mPrep CapsulesMicroscopy and Microanalysis, 2014
- Cryoplaning Technique for Visualizing the Distribution of Water in Woody Tissues by Cryoscanning Electron Microscopy2007
- Cryo-Fracturing and Cryo-Planing for In-Lens Cryo-SEM, Using a Newly Designed Diamond KnifeMicroscopy and Microanalysis, 2003
- Structure of the Developing Pea Seed Coat and the Post-phloem Transport Pathway of NutrientsAnnals of Botany, 2003
- Cryo‐planing for cryo‐scanning electron microscopyScanning, 1999