Coronary MR Angiography at 3.0 T versus That at 1.5 T: Initial Results in Patients Suspected of Having Coronary Artery Disease
Authors:
- Torsten SommerTSTorsten Sommer
- Matthias HackenbrochMHMatthias Hackenbroch
- Ulrich HoferUHUlrich Hofer
- Alexandra SchmiedelASAlexandra Schmiedel
- Winfried A. WillinekWWWinfried A. Willinek
- Sebastian FlackeSFSebastian Flacke
- Jürgen GiesekeJGJürgen Gieseke
- Frank TräberFTFrank Träber
- Rolf FimmersRFRolf Fimmers
- Harold LittHLHarold Litt
- Hans SchildHSHans Schild
- 1 March 2005
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 234 (3) , 718-725
Abstract
To prospectively evaluate the feasibility, image quality, and accuracy of coronary magnetic resonance (MR) angiography at 3.0 T in patients suspected of having coronary artery disease and to prospectively compare these results with those of coronary MR angiography performed at 1.5 T. The study was approved by the institutional review board, and informed consent was obtained from all patients. Eighteen patients (11 men, seven women; mean age, 63 years; age range, 45-76 years) suspected of having coronary artery disease who were scheduled to undergo elective conventional coronary angiography (reference standard) were included. For coronary MR angiography at 3.0 and 1.5 T, a vector electrocardiographically gated three-dimensional segmented k-space gradient-echo imaging sequence was combined with real-time respiratory navigator gating and tracking. Signal-to-noise ratios (SNRs), contrast-to-noise ratios (CNRs), scores of image quality and sensitivity and specificity for the detection of coronary artery stenosis on a segment-by-segment basis were assessed at 3.0 and 1.5 T. Data were analyzed for statistical differences by using the Wilcoxon matched-pairs test and the McNemar test. The average increase in SNR at 3.0 T with respect to that at 1.5 T was 29.5% for the left coronary artery (LCA) and 31.2% for the right coronary artery (RCA) (P < .001), and the average increase in CNR was 21.8% for the LCA and 23.5% for the RCA (P < .001). Scores of image quality (P = .77) and diagnostic accuracy for the detection of coronary artery stenoses (sensitivity and specificity: 82% and 89%, respectively, at 3.0 T vs 82% and 88% at 1.5 T; P > .99) were identical or almost identical at both field strengths. Coronary MR angiography at 3.0 T is feasible in patients suspected of having coronary artery disease and yields significant increases in SNR and CNR, although current techniques do not result in significantly improved image quality and diagnostic accuracy compared with the quality and accuracy at 1.5 T. (c) RSNA, 2005.Keywords
This publication has 22 references indexed in Scilit:
- Preliminary report on in vivo coronary MRA at 3 Tesla in humansMagnetic Resonance in Medicine, 2002
- Cardiac imaging at 4 TeslaMagnetic Resonance in Medicine, 2001
- Human cardiac imaging at 3 T using phased array coilsMagnetic Resonance in Medicine, 2000
- A generalized k-sampling scheme for 3D fast spin echoJournal of Magnetic Resonance Imaging, 2000
- Submillimeter Three-dimensional Coronary MR Angiography with Real-time Navigator Correction: Comparison of Navigator LocationsRadiology, 1999
- Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitionsMagnetic Resonance in Medicine, 1999
- Improved Coronary Artery Definition With T2-Weighted, Free-Breathing, Three-Dimensional Coronary MRACirculation, 1999
- The Intrinsic Signal-to-Noise Ratio in Human Cardiac Imaging at 1.5, 3, and 4 TJournal of Magnetic Resonance, 1997
- Broadband nuclear magnetic resonance pulses with two-dimensional spatial selectivityJournal of Applied Physics, 1989
- The intrinsic signal‐to‐noise ratio in NMR imagingMagnetic Resonance in Medicine, 1986