Abstract
This invited paper demonstrated an ultra-highly sensitive ammonia (NH3) sensor based on the light-induced thermoelastic spectroscopy (LITES) technique for the first time. A quartz tuning fork (QTF) with a resonance frequency of 32.768 kHz was employed as a detector. A fiber-coupled, continuous wave (CW), distributed feedback (DFB) diode laser emitting at 1530.33 nm was chosen as the excitation source. Wavelength modulation spectroscopy (WMS) and second-harmonic (2f) detection techniques were applied to reduce the background noise. In a one scan period, a 2f signal of the two absorption lines located at 6534.6 cm−1 and 6533.4 cm−1 were acquired simultaneously. The 2f signal amplitude at the two absorption lines was proved to be proportional to the concentration, respectively, by changing the concentration of NH3 in the analyte. The calculated R-square values of the linear fit are equal to ~0.99. The wavelength modulation depth was optimized to be 13.38 mA, and a minimum detection limit (MDL) of ~5.85 ppm was achieved for the reported NH3 sensor.
Funding Information
  • National Natural Science Foundation of China (62022032, 61875047, 61505041)
  • Natural Science Foundation of Heilongjiang Province (YQ2019F006)
  • Fundamental Research Funds for the Central Universities (none)
  • Financial Grant from the Heilongjiang Province Postdoctoral Foundation (LBH-Q18052)

This publication has 32 references indexed in Scilit: