A New Device for Assessing Ankle Dorsiflexion Motion: Reliability and Validity

Authors:

Abstract
Study Design: Clinical measurement. Objective: To determine the validity and reli- ability of measures obtained using a custom-made device for assessing ankle dorsiflexion motion and stiffness. Background: Limited dorsiflexion has been implicated in the evolution of foot pain in a number of clinical populations. Assessment of ankle dorsiflexion range of motion (ROM) is, therefore, commonly performed as part of a foot and ankle examination. Conventional goniometric assessment methods have demonstrated limited intertester reliability, while alternative methods of measurements are generally more difficult to use. The Iowa ankle range of motion (IAR0M) device was designed in an attempt to develop a simple, clinically relevant, and time- and cost-effective tool to measure ankle dorsiflexion range of motion and stiffness. Methods: Validity and intertester reliability of dorsiflexion range-of-motion measures using the IAR0M device were assessed at 10,15,20, and 25 Nm of passively applied dorsiflexion torque, with both the knee extended and flexed approxi- mately 20°. Stiffness (change in torque/change in dorsiflexion angle) values were determined using the angular change obtained between the 15- and 25-Nm torque levels. Convergent validity (n = 12) was assessed through comparison of ankle dorsiflexion angles measured simultaneously with the IAR0M device and an optoelectronic motion analysis system. Intertester reliability (n = 17) was assessed by 2 testers who took measurements within the same day. Results: Validity testing demonstrated excellent agreement (intraclass correlation coefficient [ICC] values ranging from 0.95 to 0.98). Reliability testing demonstrated good to excellent intertester agreement (ICC values ranging from 0.90 to 0.95). The ICCs for ankle joint dorsiflexion stiffness were .71 and .85 for the knee in an extended and flexed position, respectively. Conclusion: The IAROM device provides valid and reliable measurement of ankle dorsiflexion ROM. The IAR0M device also allows calculation of stiffness by measuring ROM at multiple torque levels, although the reliability of the measurement is not optimal. J Orthop Sports Phys Ther 2011;41(4):274–280. doi:10.2519/jospt.2011.3397

This publication has 24 references indexed in Scilit: