The Argo: A 65,536 channel recording system for high density neural recording in vivo
Authors:
- Kunal SahasrabuddheKSKunal Sahasrabuddhe
- Aamir A. KhanAKAamir A. Khan
- Aditya P. SinghASAditya P. Singh
- Tyler M. SternTSTyler M. Stern
- Yeena NgYNYeena Ng
- Aleksandar TadićATAleksandar Tadić
- Peter OrelPOPeter Orel
- Chris LaReauCLChris LaReau
- Daniel PouzznerDPDaniel Pouzzner
- Kurtis NishimuraKNKurtis Nishimura
- Kevin M. BoergensKBKevin M. Boergens
Preprint
Abstract
Here we demonstrate the Argo System, a massively parallel neural recording system based on platinum-iridium microwire electrode arrays bonded to a CMOS voltage amplifier array. The Argo system is the highest channel count in vivo neural recording system built to date, supporting simultaneous recording from 65,536 channels, sampled at over 32 kHz and 12-bit resolution. This system is designed for cortical recordings, compatible with both penetrating and surface microelectrodes. We have validated this system by recording spiking activity from 791 neurons in rats and cortical surface Local Field Potential (LFP) activity from over 30,000 channels in sheep. While currently adapted for head-fixed recording, the microwire-CMOS architecture is well suited for clinical translation. Thus, this demonstration helps pave the way for a future high data rate intracortical implant.Keywords
This publication has 67 references indexed in Scilit:
- A modular high-density μECoG system on macaque vlPFC for auditory cognitive decodingJournal of Neural Engineering, 2020
- High-Density, Long-Lasting, and Multi-region Electrophysiological Recordings Using Polymer Electrode ArraysNeuron, 2019
- FEF inactivation with improved optogenetic methodsProceedings of the National Academy of Sciences, 2016
- Insulation lifetime improvement of polyimide thin film neural implantsJournal of Neural Engineering, 2015
- The Emergence of Single Neurons in Clinical NeurologyNeuron, 2015
- Ultrasensitive fluorescent proteins for imaging neuronal activityNature, 2013
- The cortical angiome: an interconnected vascular network with noncolumnar patterns of blood flowNature Neuroscience, 2013
- Method of electrochemical etching of tungsten tips with controllable profilesReview of Scientific Instruments, 2012
- Neurotrophic electrode: Method of assembly and implantation into human motor speech cortexJournal of Neuroscience Methods, 2008
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008