Comparison of Right– and Left–Sided Pectoral Implantation Parameters with the Jewel Active Can Cardiodefibrillator
Authors:
- 1 February 1998
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 21 (2) , 447-451
Abstract
A total of 1,207 patients received a Medtronic jewel active, can ICD (models 721BC, 7219C), with a Transvene had in 97 centers in Europe and North America. Nineteen implants were from the right pectoral region. Patients with right–sided ICDs did not differ in terms of mean age. % male, left ventricular ejection fraction, New York Heart Association Functional Class, antiar–rhythmic drug therapy, indication for the implantable cardioverter defihrillator, and R wave values at implantation, but tended to have slightly higher pacing thresholds (1.2 ± 0.5 V vs 1.0 ± 0.6 V, P = 0.012) and higher defibrillation thresholds (14.7 ± 6.4 J vs 11.5 ± 6 J, P = 0.11) compared with patients with left sided implants. Patients with right–sided implants had a longer implantation time compared with patients with left–sided implants (118 ± 70 minutes vs 91 ± 46 minutes, P = 0.074). In follow–up, 5 patients with right–sided implantation received successful therapy for either ventricular fibrillation, (8 episodes) or ventricular tachycardia (5 episodes). No ineffective therapy from the device was delivered in any patients with right–sided implantation. Right–sided pectoral implants are feasible with the Medtronic Jewel active can ICD.Keywords
This publication has 9 references indexed in Scilit:
- Optimal electrode position for transvenous defibrillation: A prospective randomized studyJACC, 1996
- Elevated Defibrillation Threshold When Right‐Sided Venous Access is Used for Nonthoracotomy Implantable Defibrillator Lead ImplantationJournal of Cardiovascular Electrophysiology, 1995
- A simplified, single-lead unipolar transvenous cardioversion-defibrillation system.Circulation, 1993
- Clinical characteristics and outcome of patients with high defibrillation thresholds. A multicenter study.Circulation, 1992
- Prospective comparison of biphasic and monophasic shocks for implantable cardioverter-defibrillators using endocardial leadsThe American Journal of Cardiology, 1992
- Cardiac potential and potential gradient fields generated by single, combined, and sequential shocks during ventricular defibrillation.Circulation, 1992
- A prospective randomized evaluation of biphasic versus monophasic waveform pulses on defibrillation efficacy in humansJACC, 1989
- Long-term outcome with the automatic implantable cardioverter-defibrillatorJACC, 1989
- The automatic implantable cardioverter defibrillator: Efficacy, complications and survival in patients with malignant ventricular arrhythmiasJACC, 1988