Peripheral refraction in pseudophakic eyes measured by infrared scanning photoretinoscopy

Authors:
  • AOArne Ohlendorf
    Arne Ohlendorf
  • DFDominik M. Fischer
    Dominik M. Fischer
  • ABAnna R. Bruckmann
    Anna R. Bruckmann
  • USUlrich Schiefer
    Ulrich Schiefer
  • FSFrank Schaeffel
    Frank Schaeffel

Abstract
Subjects were measured with a purpose-built scanning photorefractor. The instrument was improved over previous versions. It permits measurement of semicontinuous peripheral profiles over the central 90-degree field of the retina at a faster speed (4 s/scan). RESULTS: Twenty-four pseudophakic and 43 phakic subjects were enrolled. The intraocular lenses (IOLs) induced a mean myopic shift of 2.00 diopters (D) at ±45 degrees of eccentricity in the vertical pupil meridian. Ray-tracing simulations with phakic eye and pseudophakic eye models agreed well with the experimental data. They showed that changes induced by IOLs were a consequence of an increase in astigmatism with eccentricity and a myopic shift in the spherical equivalent. CONCLUSIONS: The peripheral refractions in pseudophakic eyes were more myopic than in phakic eyes as a consequence of the optical design of the IOLs. Whether a more myopic refraction of approximately 2.00 D at 45 degrees has significant effects on visual performance must be tested. Perhaps there is room for improvement in the peripheral optics of IOLs. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned....