Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator
Authors:
Open Access
- 3 June 2024
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 47 (7) , 885-892
Abstract
Background: Electromagnetic interference (EMI) encompasses electromagnetic field signals that can be detected by a device's circuitry, potentially resulting in adverse effects such as inaccurate sensing, pacing, device mode switching, and defibrillation. EMI may impact the functioning of Cardiac Implantable Electronic Devices (CIEDs) and lead to inappropriate therapy.Method: An experimental measuring device, a loop antenna mimicking the implantable cardioverted defibrillator (ICD) antenna, was developed, and validated at the US Food and Drug Administration (FDA) and sent to Wright State University for testing. Two sets of measurements were conducted while the vehicle was connected to a 220‐Volt outlet with charging at ON and OFF. Each measurement set involved three readings at various locations, with the antenna oriented in three different positions to account for diverse patient postures. The experiment utilized a Tesla Model 3 electric vehicle (EV), assessing scenarios both inside and outside the car, including the driver's seat, driver's seat floor, passenger's seat, rear seat, rear seat floor, cup holder, charging port (car), and near the charging station.Results: The detected voltage (max 400 to 504 millivolts) around the cup holder inside the car differed from all other measurement scenarios.Conclusion: The investigation highlights the identification of EMI signals originating from an EV) that could potentially interrupt the functionality of a Subcutaneous Implantable Cardioverter‐Defibrillator (S‐ICD). These signals fell within the R‐wave Spectrum of 30–300 Hz. Further in‐vivo studies are essential to determine accurately the level of interference between S‐ICDs and EMI from Electric Vehicles.Keywords
This publication has 15 references indexed in Scilit:
- Electric Vehicles Under Low Temperatures: A Review on Battery Performance, Charging Needs, and Power Grid ImpactsIEEE Access, 2023
- 2022 ESC Guidelines on cardiovascular assessment and management of patients undergoing non-cardiac surgeryEuropean Heart Journal, 2022
- Spotlight on S-ICD™ therapy: 10 years of clinical experience and innovationEP Europace, 2019
- Subcutaneous Versus Transvenous Implantable Defibrillator TherapyJACC: Clinical Electrophysiology, 2017
- Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD RegistryEuropean Heart Journal, 2014
- Electromagnetic Interference and Implanted Cardiac Devices: The Nonmedical Environment (Part I)Clinical Cardiology, 2012
- Pacemaker sensitivity to 50 Hz noise voltagesEP Europace, 2011
- In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillatorsHeart Rhythm, 2010
- Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable defibrillatorEP Europace, 2006
- Case Report: Inappropriate Implantable Cardioverter Defibrillator Discharge from Sensing of External Alternating Current LeakJournal of Interventional Cardiac Electrophysiology, 2002