Modeling scintillator‐photodiodes as detectors for megavoltage CT

Authors:
  • SSS. Steciw
    S. Steciw
  • BFB. G. Fallone
    B. G. Fallone
  • SRS. Rathee
    S. Rathee

Abstract
The use of cadmium tungstateand cesium iodide [CsI(Tl)] scintillation detectors is studied in megavoltage computed tomography (MVCT). A model describing the signal acquired from a scintillation detector has been developed which contains two steps: (1) the calculation of the energy deposited in the crystal due to MeV photons using theEGSnrcMonte Carlo code; and (2) the transport of the optical photons generated in the crystal voxels to photodiodes using the optical Monte Carlo codeDETECT2000. The measured detector signals in singleand CsI(Tl) scintillation crystals of baseand heights 0.4, 1, 1.2, 1.6 and 2 cm were, generally, in good agreement with the signals calculated with the model. A prototype detector array which contains 8crystals, eachin contact with a 16‐element array of photodiodes was built. The measured attenuation of a Cobalt‐60 beam as a function of solid water thickness behaves linearly. The frequency dependent modulation transfer functionnoise power spectrumand detective quantum efficiencywere measured for 1.25 MeV photons (in a Cobalt‐60 beam). For 6 MV photons, only thewas measured from a linear accelerator, where large pulse‐to‐pulse fluctuations in the output of the linear accelerator did not allow the measurement of theA two‐step Monte Carlo simulation was used to model the detector's andThe DQE(0) of the detector array was found to be 26% and 19% for 1.25 MeV and 6 MV photons, respectively. For 1.25 MeV photons, the maximum discrepancies between the measured and modeledrelativeand thewere found to be 1.5%, 1.2%, and 1.9%, respectively. For the 6 MV beam, the maximum discrepancy between the modeled and the measuredwas found to be 2.5%. The modeling is sufficiently accurate for designing appropriate detectors for MVCT.

This publication has 31 references indexed in Scilit: