The Peregrine Ion Trap Mass Spectrometer (PITMS) is a mass spectrometer instrument that operated during the Astrobotic Peregrine Mission-1 as part of the NASA Commercial Lunar Payload Services initiative. This paper describes the instrument and investigation design, development, and planning conducted by the PITMS team, consisting of a successful partnership between NASA Goddard Space Flight Center (GSFC), The Open University, NASA, and ESA. PITMS was designed to measure the abundance and temporal variability of volatile species in the near-surface lunar exosphere from a landed platform on the lunar surface. The PITMS instrument consisted of a European Space Agency–provided Exospheric Mass Spectrometer (including sensor, electronics, controller, and power supply boards) and a GSFC wrapper that provided structural elements, thermal control, and a deployable dust cover. PITMS was designed to operate as a passive sampler, where ambient gases would enter PITMS through an aperture, diffuse around the mass analyzer cavity, become ionized by electron impact and trapped in an RF field, and then sequentially be released to a detector to build a mass spectrum. PITMS was capable of measuring species with a mass-to-charge ratio (m/z) from 10 to 150 Da, with a mass resolution of approximately 0.5 amu. The PITMS science investigation was planned to be operated by GSFC with an international team of scientists. Though the mission did not achieve its lunar landing, information about the PITMS instrument and planning is provided to be able to understand and effectively use data that will be forthcoming from the investigation.Keywords
Funding Information
- National Aeronautics and Space Administration (364688.05.12.01)
This publication has 28 references indexed in Scilit:
- Volatile interactions with the lunar surfaceGeochemistry, 2022
- Particles and Photons as Drivers for Particle Release from the Surfaces of the Moon and MercurySpace Science Reviews, 2022
- Contributions of solar wind and micrometeoroids to molecular hydrogen in the lunar exosphereIcarus, 2017
- Methane in the lunar exosphere: Implications for solar wind carbon escapeGeophysical Research Letters, 2016
- Temperature programmed desorption studies of water interactions with Apollo lunar samples 12001 and 72501Icarus, 2015
- Variability of helium, neon, and argon in the lunar exosphere as observed by the LADEE NMS instrumentGeophysical Research Letters, 2015
- Spillage of lunar polar crater volatiles onto adjacent terrains: The case for dynamic processesGeophysical Research Letters, 2015
- Thermal stability of water and hydroxyl on the surface of the Moon from temperature-programmed desorption measurements of lunar analog materialsIcarus, 2011
- Ptolemy – an Instrument to Measure Stable Isotopic Ratios of Key Volatiles on a Cometary NucleusSpace Science Reviews, 2006