Hollow Implants Retrieved for Fracture: A Light and Scanning Electron Microscope Analysis of 4 Cases
- 1 February 1998
- journal article
- research article
- Published by Wiley in Journal of Periodontology
- Vol. 69 (2) , 185-189
Abstract
One of the possible complications of implant treatment is the occurrence of an implant fracture. Metal fatigue and biomechanical overload seem to be the most common causes of fractured implants. This study evaluated 4 implants (3 hollow cylinders and 1 hollow screw) which fractured after a mean loading period of 2.8 years. All implants had a 4 mm diameter and had been inserted in a posterior location. In 3 cases parafunctional habits were present. In all cases a vertical resorption of the peri‐implant bone was present. The endosseous portion of the implant presented always a very high bone‐implant contact percentage. Scanning electron microscopic examination showed that at least one of the implant holes was involved in the fracture line; no porosities or material defects were observed on the fractured surface of the implant. In hollow implants the holes could represent a site of less resistance. J Periodontol 1998;69:185–189.Keywords
This publication has 5 references indexed in Scilit:
- High-precision, cost-effective cutting system for producing thin sections of oral tissues containing dental implantsBiomaterials, 1997
- Fractures of Hydroxyapatite‐Coated Blade Implants Connected With Natural Teeth. A Histological Study Using SEM, Light Microscopy, and an Image Processing SystemJournal of Periodontology, 1996
- LOAD DISTRIBUTION ON IMPLANTS WITH A CANTILEVERED SUPERSTRUCTUREImplant Dentistry, 1996
- On clinical loading of osseointegrated implants. A methodological and clinical study.Clinical Oral Implants Research, 1993
- Marginal tissue reactions at osseointegrated titanium fixturesInternational Journal of Oral and Maxillofacial Surgery, 1986