Detection of counterfeit tablets of an antiviral drug using δ34S measurements by MC-ICP-MS and confirmation by LA-MC-ICP-MS and HPLC-MC-ICP-MS
Authors:
- 16 July 2008
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 23 (9) , 1294-1299
Abstract
A new approach for pharmaceutical counterfeit detection using laser ablation multicollector inductively coupled plasma mass spectrometry (LA-MC-ICP-MS) and high performance liquid chromatography multicollector inductively coupled plasma mass spectrometry (HPLC-MC-ICP-MS) has been developed. Laser ablation and chromatographic separation parameters have been previously optimised and helium has been used as the carrier gas for better sensitivity. A homogeneity study of 288 pharmaceutical tablets from different batches of the genuine drug has been performed by MC-ICP-MS and characteristic sulfur isotopic signature has been obtained for the genuine product. δ34S measurements by MC-ICP-MS using silicon internal standardisation for the correction of instrumental mass bias effects led to a δ34S of 3.6‰ and an associated combined expanded uncertainty of 1‰ (k = 2). The active pharmaceutical ingredient (API) of the tablets, a S-containing compound, has been separated by HPLC and sulfur isotope amount ratios have been measured by MC-ICP-MS. δ34S values obtained by HPLC-MC-ICP-MS for the genuine tablets agreed with those obtained by laser ablation MC-ICP-MS, thus confirming that the sulfur isotopic signature is inherent to the S-containing API. Following the initial development work, a blind exercise was performed for four hundred tablets. The discriminating power of the technique was assessed and uncertainties associated to δ34S values for counterfeit and genuine tablets varied depending on the sample introduction technique utilised. The three approaches were able to differentiate genuine from counterfeit tablets. During the exercise, four distinct groups of counterfeit tablets were detected. The MC-ICP-MS method therefore has potential as a rapid screening tool for pharmaceutical counterfeit detection and classification.Keywords
This publication has 29 references indexed in Scilit:
- Systematic comparison of δ 34 S measurements by multicollector inductively coupled plasma mass spectrometry and evaluation of full uncertainty budget using two different metrological approachesRapid Communications in Mass Spectrometry, 2008
- Measurement of strontium isotope ratios by MC-ICP-MS after on-line Rb–Sr ion chromatography separationJournal of Analytical Atomic Spectrometry, 2007
- δ34S Measurements of Sulfur by Multicollector Inductively Coupled Plasma Mass SpectrometryAnalytical Chemistry, 2006
- Screening suspected counterfeit Viagra® and imitations of Viagra® with near-infrared spectroscopyJournal of Pharmaceutical and Biomedical Analysis, 2006
- Measurement of sulfur isotope composition ( δ 34 S) by multiple‐collector thermal ionization mass spectrometry using a 33 S‐ 36 S double spikeRapid Communications in Mass Spectrometry, 2005
- Stable isotopic characterization of active pharmaceutical ingredientsJournal of Pharmaceutical and Biomedical Analysis, 2004
- Isotope ratio determination of antimony from the transient signal of trimethylstibine by GC-MC-ICP-MS and GC-ICP-TOF-MSJournal of Analytical Atomic Spectrometry, 2003
- High accuracy analysis of low level sulfur in diesel fuel by isotope dilution high resolution ICP-MS, using silicon for mass bias correction of natural isotope ratiosJournal of Analytical Atomic Spectrometry, 2001
- Precise sulfur isotope ratio measurements in trace concentration of sulfur by inductively coupled plasma double focusing sector field mass spectrometryJournal of Analytical Atomic Spectrometry, 1999
- Enhanced sensitivity in laser ablation-ICP mass spectrometry using helium-argon mixtures as aerosol carrierJournal of Analytical Atomic Spectrometry, 1999