MR-guided Vacuum-assisted Breast Biopsy: Accuracy of Targeting and Success in Sampling in a Phantom Model
Authors:
- Constance D. LehmanCLConstance D. Lehman
- Taro AikawaTATaro Aikawa
- 1 September 2004
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 232 (3) , 911-914
Abstract
An 11-gauge MR-compatible system was designed for use in magnetic resonance vacuum-assisted breast biopsy. The system uses a detachable needle with minimal artifact to allow imaging after placement and before biopsy to confirm lesion location. A phantom study involving 16 biopsies of lesions smaller than 10 mm was conducted to assess the performance of this system. Fifteen (94%) of 16 biopsies resulted in successful lesion sampling. Eighty-five percent of all core samples contained specimen materials targeted lesion material. Further research to evaluate the efficiency, accuracy, and safety of this system in a patient population is recommended.Keywords
This publication has 7 references indexed in Scilit:
- Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breastEuropean Radiology, 2002
- MR-guided interventional breast procedures considering vacuum biopsy in particularEuropean Journal of Radiology, 2002
- Interventional MRI of the breast: lesion localisation and biopsyEuropean Radiology, 2000
- MR-guided percutaneous excisional and incisional biopsy of breast lesionsEuropean Radiology, 1999
- Comparison of rebiopsy rates after stereotactic core needle biopsy of the breast with 11-gauge vacuum suction probe versus 14-gauge needle and automatic gun.American Journal of Roentgenology, 1999
- Evaluation of 14- and 11-gauge directional, vacuum-assisted biopsy probes and 14-gauge biopsy guns in a breast parenchymal model.Radiology, 1997
- Stereotactic breast biopsy of atypical ductal hyperplasia and ductal carcinoma in situ lesions: improved accuracy with directional, vacuum-assisted biopsy.Radiology, 1997