Abstract
Objectives: (1) To determine ciliary beat frequency (CBF) using a consumer‐grade cellphone camera and MATLAB and (2) to evaluate the effectiveness and accuracy of the proposed method.Study Design: Prospective animal study.Setting: Academic otolaryngology department research laboratory.Methods: Five ex vivo tracheal samples were extracted from 3 freshly euthanized (P <. 001). The sample with 10% formalin did not display any data beyond DC noise. Compared with 240 fps, the means of other frame rates/methods (120, 60, 30 fps; manual counting) at 23°C all showed no statistical difference (P >. 05).Conclusion: There is no significant difference between CBF measured via visual inspection and that analyzed by the developed program. Furthermore, all tested acquisition rates are shown to be effective, providing a fast and inexpensive alternative to current CBF measurement protocols.