Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients
Authors:
Open Access
- 1 January 2020
- journal article
- research article
- Published by AME Publishing Company in Translational Cancer Research
- Vol. 9 (1) , 166-172
Abstract
Background: To evaluate the dissociation effect of optical fluorescence platelet counting of BC-6800 hematology analyzer on ethylene diamine tetraacetic acid-dependent pseudo thrombocytopenia (EDTA-PTCP) samples. Methods: Twenty-three finally identified EDTA-PTCP samples were recruited in this study using criteria as follow: (I) impedance platelet counts lower than 100×109/L with instrument “platelet aggregation” flag; (II) existence of platelet clumps in the blood smear; (III) obviously higher platelet counts without “platelet aggregation” flag and no platelet clumps in blood smear after repeating phlebotomy using citrate anticoagulated tubes. The BC-6800 hematology analyzer and the XE-2100 hematology analyzer were used to test 23 EDTA-PTCP samples and 30 controls on both the impedance channel and the reticulocyte channel. The dissociation rate was defined as optical fluorescence platelet counts in the EDTA tubes/impedance platelet counts in citrate tubes ×100%. Results: BC-6800 analyzer’s optical fluorescence platelet counts of EDTA-PTCP samples were significantly higher than impedance platelet counts (t=4.33, P=0.00) and comparable with the platelet counts of re-collected samples in tubes containing citrate anticoagulant. On BC-6800 hematology analyzer, 22 of 23 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 93%. On the XE-2100 hematology analyzer, 1 of the 17 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 56%. Conclusions: Optical fluorescence platelet counting of BC-6800 Hematology Analyzer is effective for the correction of spurious low platelet counts in EDTA-PTCP patients, and its dissociation effect on EDTA-PTCP samples was independent of fluorescent dye staining.Keywords
This publication has 18 references indexed in Scilit:
- Discovery and Correction of Spurious Low Platelet Counts due to EDTA‐Dependent PseudothrombocytopeniaJournal of Clinical Laboratory Analysis, 2014
- Pseudothrombocytopenia in a Child with DengueIndian Journal of Pediatrics, 2014
- Pseudothrombocytopenia in a patient receiving romiplostim for immune thrombocytopenic purpuraAnnals of Hematology, 2013
- Effective estimation of correct platelet counts in pseudothrombocytopenia using an alternative anticoagulant based on magnesium saltBritish Journal of Haematology, 2013
- Platelet Counting2013
- Ethylenediaminetetraacetic acid-dependent pseudothrombocytopenia associated with neuroendocrine carcinoma: A case reportOncology Letters, 2012
- EDTA-dependent pseudothrombocytopenia: further insights and recommendations for prevention of a clinically threatening artifactClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Novel method to dissociate platelet clumps in EDTA-dependent pseudothrombocytopenia based on the pathophysiological mechanismClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Transient appearance of postoperative EDTA-dependent pseudothrombocytopenia in a patient after gastrectomyPlatelets, 2010
- Spurious counts and spurious results on haematology analysers: a review. Part I: plateletsInternational Journal of Laboratory Hematology, 2007