The scanning laser ophthalmoscope
Authors:
- 1 May 1997
- journal article
- review article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 42 (5) , 951-966
Abstract
The imaging of the fundus of the eye poses two major technical challenges. First, it is necessary for both the illuminating and reflected beams to pass through the same aperture, the iris. In some commonly used instruments this leads to the use of levels of illumination close to the maximum tolerable by a patient. Second, in order to visualize the different structures present in the various layers of the fundus it is necessary to perform tomographic imaging. The scanning laser ophthalmoscope provides an answer to these particular problems. By scanning the fundus with a narrow laser beam most of the area of the iris is then available for the reflected light and so the intensity of the illuminating beam can be kept low, making it more acceptable for patients. The use of confocal imaging allows 3D images to be produced. In this short review the performance of the instrument will be discussed and its application to a number of clinical problems in ophthalmology considered. Finally there will be a brief description of other instrumentation currently under development.Keywords
This publication has 69 references indexed in Scilit:
- An Image-Processing Strategy for the Segmentation and Quantification of Microaneurysms in Fluorescein Angiograms of the Ocular FundusComputers and Biomedical Research, 1996
- A new fluorescent imaging procedurein vivofor evaluation of the retinal microcirculation in ratsCurrent Eye Research, 1995
- SIMULTANEOUS INDOCYANINE GREEN AND FLUORESCEIN ANGIOGRAPHYRetina, 1995
- Clinical investigation of an infrared digital scanning laser ophthalmoscope.British Journal of Ophthalmology, 1994
- Reproducibility of topometric data acquisition in normal and glaucomatous optic nerve heads with the laser tomographic scannerGraefe's Archive for Clinical and Experimental Ophthalmology, 1993
- Automated detection and quantification of retinal exudatesGraefe's Archive for Clinical and Experimental Ophthalmology, 1993
- Use of temporal information to quantify vascular leakage in fluorescein angiography of the retinaClinical Physics and Physiological Measurement, 1990
- Active optical depth resolution improvement of the laser tomographic scannerApplied Optics, 1989
- Digital laser scanning fundus cameraApplied Optics, 1987
- Flying spot TV ophthalmoscopeApplied Optics, 1980