Classification Scheme of Heating Risk during MRI Scans on Patients with Orthopaedic Prostheses
Authors:
- Valeria ClementiVCValeria Clementi
- Umberto ZanovelloUZUmberto Zanovello
- Alessandro ArduinoAAAlessandro Arduino
- Cristina AncaraniCACristina Ancarani
- Fabio BaruffaldiFBFabio Baruffaldi
- Barbara BordiniBBBarbara Bordini
- Mario ChiampiMCMario Chiampi
- Luca ZilbertiLZLuca Zilberti
- Oriano BottauscioOBOriano Bottauscio
Open Access
- 2 August 2022
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 12 (8) , 1873
Abstract
Due to the large variety of possible clinical scenarios, a reliable heating-risk assessment is not straightforward when patients with arthroplasty undergo MRI scans. This paper proposes a simple procedure to estimate the thermal effects induced in patients with hip, knee, or shoulder arthroplasty during MRI exams. The most representative clinical scenarios were identified by a preliminary frequency analysis, based on clinical service databases, collecting MRI exams of 11,658 implant carrier patients. The thermal effects produced by radiofrequency and switching gradient fields were investigated through 588 numerical simulations performed on an ASTM-like phantom, considering four prostheses, two static field values, seven MR sequences, and seven regions of imaging. The risk assessment was inspired by standards for radiofrequency fields and by scientific studies for gradient fields. Three risk tiers were defined for the radiofrequency, in terms of whole-body and local SAR averages, and for GC fields, in terms of temperature elevation. Only 50 out of 588 scenarios require some caution to be managed. Results showed that the whole-body SAR is not a self-reliant safety parameter for patients with metallic implants. The proposed numerical procedure can be easily extended to any other scenario, including the use of detailed anatomical models.Keywords
Funding Information
- European Association of National Metrology Institutes (17IND01 MIMAS)
- European Association of National Metrology Institutes (17IND01 MIMAS)
This publication has 22 references indexed in Scilit:
- Douglas–Gunn Method Applied to Dosimetric Assessment in Magnetic Resonance ImagingIEEE Transactions on Magnetics, 2017
- Gradient heating of bulk metallic implants can be a safety concern in MRIMagnetic Resonance in Medicine, 2017
- Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective studyMagnetic Resonance Materials in Physics, Biology and Medicine, 2015
- Numerical prediction of temperature elevation induced around metallic hip prostheses by traditional, split, and uniplanar gradient coilsMagnetic Resonance in Medicine, 2015
- Clinical Implementation of MRI of Joint ArthroplastyAmerican Journal of Roentgenology, 2014
- Computational and experimental studies of an orthopedic implant: MRI‐related heating at 1.5‐T/64‐MHz and 3‐T/128‐MHzJournal of Magnetic Resonance Imaging, 2012
- Specific absorption rate intersubject variability in 7T parallel transmit MRI of the headMagnetic Resonance in Medicine, 2012
- A specific absorption rate prediction concept for parallel transmission MRMagnetic Resonance in Medicine, 2012
- Numerical simulation of SAR induced around Co‐Cr‐Mo hip prostheses in situ exposed to RF fields associated with 1.5 and 3 T MRI body coilsMagnetic Resonance in Medicine, 2011