A modified transosseous quadriceps tendon refixation technique demonstrates advantageous primary stability: A biomechanical investigation
Open Access
- 1 January 2026
- journal article
- research article
- Published by Wiley in Journal of Experimental Orthopaedics
- Vol. 13 (1) , e70695
Abstract
Purpose: Recent biomechanical research suggests suture anchor (SA) refixation may be superior to conventional transosseous longitudinal (TL) repair for quadriceps tendon ruptures. This study aimed to evaluate whether a modified transosseous technique with anteriorly angled drill holes enhances primary stability compared to the aforementioned techniques. Methods: Twenty‐one human cadaveric patellae were randomly assigned to three groups ( n = 7): SA (two SAs), TL (three longitudinal transosseous drill holes) and TA (three anteriorly angled 45° transosseous drill holes). Identical suture material (Ethibond #2) and a four‐strand suture configuration were used in all groups. Prior to fixation, computed tomography (CT) scans measured bone mineral density (BMD) and cortical thickness. Biomechanical testing simulated early postoperative loading using a universal testing machine: preload (20 N for 30 s), cyclic loading (500 cycles, 20–100 N at 1 Hz) and a load‐to‐failure test (20 mm/s). As primary outcome, ultimate load to failure was recorded; secondarily, cyclic elongation and stiffness were measured. One‐way analysis of variance (ANOVA) with Bonferroni‐corrected post hoc t tests was used for statistical analysis. Results: The TA group had a significantly higher ultimate load to failure than the TL group (472.7 ± 22.4 vs. 344.8 ± 108.8 N, p < 0.05). No significant differences were found between the SA group (438.9 ± 77.4 N) and the other groups. The TL group showed significantly greater cyclic elongation (6.22 ± 1.03 mm) compared to the SA group (4.42 ± 0.82 mm, p < 0.01). The TA group (5.03 ± 0.73 mm) showed no significant difference compared to either group. Stiffness did not differ significantly ( p > 0.05). Conclusion: The modified transosseous technique with anterior drill holes yields higher failure loads and less elongation than the conventional TL method, with biomechanical performance comparable to SA fixation, supporting its use as a viable alternative. Level of Evidence: N/A.Keywords
This publication has 25 references indexed in Scilit:
- Editorial Commentary : Quadriceps Tendon Suture Anchor Repair Provides Better Patient‐Reported Outcomes Than Transosseous Tunnel Repair: Defining the New Gold StandardArthroscopy, 2023
- Complications and re‐operations after extensor mechanism repair surgery in a large cross‐sectional cohort: females and tobacco‐users at highest risk for adverse outcomesKnee Surgery, Sports Traumatology, Arthroscopy, 2022
- Midterm outcome and strength assessment after quadriceps tendon refixation with suture anchorsEuropean Journal of Orthopaedic Surgery & Traumatology, 2022
- Soft Tissue Fixation Strategies of Human Quadriceps Tendon Grafts: A Biomechanical StudyArthroscopy, 2019
- Knotless Suture Anchor With Suture Tape Quadriceps Tendon Repair Is Biomechanically Superior to Transosseous and Traditional Suture Anchor–Based Repairs in a Cadaveric ModelArthroscopy, 2016
- Biomechanical Evaluation of Suture Anchor Versus Transosseous Tunnel Quadriceps Tendon Repair TechniquesArthroscopy, 2016
- Gap formation in a transosseous rotator cuff repair as a function of bone qualityClinical Biomechanics, 2014
- Suture anchor repair yields better biomechanical properties than transosseous sutures in ruptured quadriceps tendonsKnee Surgery, Sports Traumatology, Arthroscopy, 2014
- The epidemiology of musculoskeletal tendinous and ligamentous injuriesInjury, 2008
- A Biomechanical and Radiographic Analysis of Standard and Intracortical Suture Anchors for Arthroscopic Rotator Cuff RepairArthroscopy, 2006