Risk of airway fire with the use of KTP laser and high flow humidified oxygen delivery in a laryngeal surgery model
Authors:
Open Access
- 11 January 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-9
Abstract
Airway surgery presents a unique environment for operating room fire to occur. This study aims to explore the factors of combustion when using KTP laser with high flow oxygen in an ex-vivo model. The variables tested were varying tissue type, tissue condition, oxygen concentration, laser setting, and smoke evacuation in a stainless-steel model. Outcome measures were time of lasing to the first spark and/or flame. A multivariate Cox proportional hazard model was used to determine the risk of spark and flame across the different risk factors. For every 10% increase in oxygen concentration above 60% the risk of flame increased by a factor of 2.3. Continuous laser setting at 2.6 W increased the risk by a factor of 72.8. The risk of lasing adipose tissue is 7.3 times higher than that of muscle. Charred tissue increases the risk of flame by a factor of 92.8. Flame occurred without a preceding spark 93.6% of the time. Using KTP laser in the pulsed mode with low wattages, minimising lasing time, reducing the oxygen concentration and avoiding lasing adipose or charred tissue produce a relatively low estimated risk of spark or flame.Keywords
All Related Versions
Funding Information
- Royal Australasian College of Surgeons (142986)
This publication has 25 references indexed in Scilit:
- Hazard of CO2 laser-induced airway fire in laryngeal surgery: experimental data of contributing factorsEuropean Archives of Oto-Rhino-Laryngology, 2013
- Practice Advisory for the Prevention and Management of Operating Room FiresAnesthesiology, 2013
- Use of Lasers in Laryngeal SurgeryJournal of Voice, 2010
- Impact of oxygen concentration and laser power on occurrence of intraluminal fires during shared-airway surgery: an investigationThe Journal of Laryngology & Otology, 2008
- Effects of 532 nm Pulsed‐KTP Laser Parameters on Vessel Ablation in the Avian Chorioallantoic Membrane: Implications for Vocal Fold MucosaThe Laryngoscope, 2007
- Reducing the Incidence of Surgical Fires: Supplying Nasal Cannulae with sub-100% O2 Gas Mixtures from Anesthesia MachinesAnesthesia & Analgesia, 2005
- Airflow through the auxiliary line of the laser fiber prevents ignition of intra-airway fire during endoscopic laser surgeryLasers in Surgery and Medicine, 2002
- Airway ignition during CO2 laser laryngeal surgery and high frequency jet ventilationEuropean Journal of Anaesthesiology, 2000
- Airway ignition during co2 laser laryngeal surgery and high frequency jet ventilationEuropean Journal of Anaesthesiology, 2000
- Airway Fire during CO2 Laser Surgery Using a Xomed Laser Endotracheal TubeAnesthesiology, 1990