Neurobiochemical changes in the vicinity of a nanostructured neural implant
Open Access
- 24 October 2016
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 6 (1) , 35944
Abstract
Neural interface technologies including recording and stimulation electrodes are currently in the early phase of clinical trials aiming to help patients with spinal cord injuries, degenerative disorders, strokes interrupting descending motor pathways, or limb amputations. Their lifetime is of key importance; however, it is limited by the foreign body response of the tissue causing the loss of neurons and a reactive astrogliosis around the implant surface. Improving the biocompatibility of implant surfaces, especially promoting neuronal attachment and regeneration is therefore essential. In our work, bioactive properties of implanted black polySi nanostructured surfaces (520–800 nm long nanopillars with a diameter of 150–200 nm) were investigated and compared to microstructured Si surfaces in eight-week-long in vivo experiments. Glial encapsulation and local neuronal cell loss were characterised using GFAP and NeuN immunostaining respectively, followed by systematic image analysis. Regarding the severity of gliosis, no significant difference was observed in the vicinity of the different implant surfaces, however, the number of surviving neurons close to the nanostructured surface was higher than that of the microstructured ones. Our results imply that the functionality of implanted microelectrodes covered by Si nanopillars may lead to improved long-term recordings.Keywords
This publication has 61 references indexed in Scilit:
- Virtual typing by people with tetraplegia using a self-calibrating intracortical brain-computer interfaceScience Translational Medicine, 2015
- Chronic In Vivo Evaluation of PEDOT/CNT for Stable Neural RecordingsIEEE Transactions on Biomedical Engineering, 2015
- BBB leakage, astrogliosis, and tissue loss correlate with silicon microelectrode array recording performanceBiomaterials, 2015
- Reach and grasp by people with tetraplegia using a neurally controlled robotic armNature, 2012
- Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortexJournal of Neural Engineering, 2011
- Neural control of cursor trajectory and click by a human with tetraplegia 1000 days after implant of an intracortical microelectrode arrayJournal of Neural Engineering, 2011
- The surface immobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interfaceBiomaterials, 2011
- Bridging the Brain to the World: A Perspective on Neural Interface SystemsNeuron, 2008
- CORTICAL NEURAL PROSTHETICSAnnual Review of Neuroscience, 2004
- Connecting cortex to machines: recent advances in brain interfacesNature Neuroscience, 2002