Abstract
Background: The aim of the present study was to examine the level of unconsciousness measured with bispectral index (BIS) at different minimal alveolar concentration (MAC) levels of sevoflurane, and to study the hemodynamic and BIS reactions during noxious stimulation with transcutaneous electrical nerve stimulation (TENS) and an ice water pain test (IWP).Methods: This study was approved by the Ethics Committee and was performed on 10 healthy, young volunteers (six males and four females), ASA physical status I. Anesthesia was induced and maintained with sevoflurane in an oxygen/air mixture. The volunteers were spontaneously breathing, but if necessary, ventilation was mechanically supported. TENS and IWP were performed at 1.0, 1.5 and 2.0 MAC of sevoflurane.Results: At 1.0 MAC, there was a significant increase in BIS during pain stimulation both with IWP (PPConclusion: BIS, heart rate and blood pressure increased during pain stimulation at 1.0 MAC but not at 1.5 MAC of sevoflurane. There was a remarkable variation in BIS at 2.0 MAC of sevoflurane, with BIS values indicating wakefulness despite clinical signs of deep anesthesia. This BIS variation is probably caused by epileptogenic activity due to sevoflurane.

This publication has 22 references indexed in Scilit: