Background The eye lens is particularly vulnerable to radiation during endovascular treatments for intracranial aneurysms due to prolonged exposure under high magnification. This study presents a lens tracking method to monitor lens positions during procedures and estimate radiation doses using a phantom. Methods A consecutive series of patients treated between January and March 2023 were retrospectively reviewed. A lens tracking method was used to mark lenses on three-dimensional source images to track their positions on two-dimensional working views. An anthropomorphic head phantom and photoluminescent glass dosimeters were used to estimate lens exposure under simulated conditions. We also evaluated potential radiation reduction with collimation based on the lens position. Results Among 42 patients, 20 (48%) had their eye lens in the field of view (FOV). In 10 of these cases, collimation could have prevented direct exposure. The phantom study showed a median lens dose of 7.32 (5.02–9.59) mGy. Lenses within the anterior–posterior (AP) plane FOV received higher doses compared with those outside it (4.32 mGy vs 1.76 mGy, PConclusion In this study, lenses were often included in the FOV, thereby receiving higher radiation doses. This underscores the importance of tracking and excluding lenses from the FOV to reduce radiation exposure during procedures.Keywords
Funding Information
- Asan Institute for Life Sciences, Asan Medical Center (2024IT0004)
This publication has 27 references indexed in Scilit:
- Impact of a selective lens dose reduction protocol in 3D rotational angiography on radiation exposure to the eye lens during cerebral angiography: a randomized controlled trialJournal of NeuroInterventional Surgery, 2023
- Measurement uncertainty analysis of radiophotoluminescent glass dosimeter reader system based on GD-352M for estimation of protection quantityNuclear Engineering and Technology, 2022
- Feasibility of low-dose digital subtraction angiography protocols for the endovascular treatment of intracranial dural arteriovenous fistulasNeuroradiology, 2020
- Radiation dose and fluoroscopy time of modern endovascular treatment techniques in patients with saccular unruptured intracranial aneurysmsEuropean Radiology, 2020
- Estimated radiation dose according to the craniocaudal angle in cerebral digital subtraction angiography: Patient and phantom studyJournal Of Neuroradiology, 2019
- Analysis of Radiation Doses and Dose Reduction Strategies During Cerebral Digital Subtraction AngiographyWorld Neurosurgery, 2017
- Real-time eye lens dose monitoring during cerebral angiography proceduresEuropean Radiology, 2015
- ICRP PUBLICATION 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection ContextAnnals of the ICRP, 2012
- Local skin and eye lens equivalent doses in interventional neuroradiologyEuropean Radiology, 2009
- Preface, Executive Summary and GlossaryAnnals of the ICRP, 2007