A SIMPLE PREPARATION OF GRAPHITE-COATED GRIDS FOR HIGH RESOLUTION ELECTRON MICROSCOPY

Authors:

Abstract
High resolution electron microscopy requires that the specimen supports be smooth. With this in mind a procedure was developed for cleaving single graphite crystals into thin flakes suitable as specimen supports. Fernfindez-Morfin (1) was the first to use graphite and mica crystals for specimen supports. He as well as Sprague (2) obtained the thin sheets of mica and graphite required through successive cleavages of single crystals by attaching adhesives or plastics to each side of the crystal and then pulling these apart. We have found the complete removal of the adhesive virtually impossible. In this note we report a method in which no adhesives are used, and smooth, clean, thin graphite fihns are readily obtained. Single crystals of graphite are obtained from marble embedded with graphite (2). This is mined in Essex County, New York, and can be purchased through Ward's Natural Science Establishment, Inc., Rochester, New York. Concentrated nitric acid dissolves the marble, leaving the graphite and a small amount of quartz. This residue is washed in tap water. A drop of clean ethanol is put on a clean, glass microscope slide. A good, unfragmented, smooth, graphite crystal is placed on the drop of ethanol. Another slide is placed over this, and the sandwich is lowered into liquid nitrogen and left there until the rapid boiling ceases. The glass-graphite sandwich is then removed and the glass slides are forced apart. It is found that the frozen ethanol

This publication has 2 references indexed in Scilit: