Estimate of radiation dose to the lens in balloon sinuplasty
- 1 December 2007
- journal article
- Published by Wiley in Otolaryngology–Head and Neck Surgery
- Vol. 137 (6) , 953-955
Abstract
Objectives: To estimate radiation dose to the lens during fluoroscopy‐assisted balloon dilation of sinus ostia and to estimate the length of fluoroscopy time that will be tolerated.Methods: Radiation dose to the lens per unit fluoroscopy time was extrapolated from published data in fluoroscopically‐guided balloon dacrocystoplasty.Results: Mean exposure varies from 0.041 (right eye) to 0.284 (left eye) mGy/sec (P = 0.037) because the left eye is closer to the radiation source during the lateral projection. Threshold for lenticular opacity has been estimated as 500 mGy, which would be attained in the left eye after approximately 29 minutes of fluoroscopy.Conclusions: The lens may tolerate <30 minutes of fluoroscopy during balloon sinuplasty. The left eye receives a potentially greater radiation dose because of its proximity to the source in the lateral projection. Dedicated studies are indicated to specifically examine this issue during balloon sinuplasty.Keywords
This publication has 5 references indexed in Scilit:
- Safety and Outcomes of Balloon Catheter Sinusotomy: A Multicenter 24‐Week Analysis in 115 PatientsOtolaryngology–Head and Neck Surgery, 2007
- Catheter-Based Dilation of the Sinus Ostia: Initial Safety and Feasibility Analysis in a Cadaver ModelAmerican Journal of Rhinology, 2006
- Radiation Dose to the Eyes and Parotids during CT of the SinusesOtolaryngology–Head and Neck Surgery, 2005
- Lens of the Eye: Radiation Dose in Balloon DacryocystoplastyRadiology, 2000
- Conversion factor f for X-ray beam qualities, specified by peak tube potential and HVL valuePhysics in Medicine & Biology, 1987