Reduced Relative Sensitivity of the Elecsys SARS-CoV-2 Antigen Assay in Saliva Compared to Nasopharyngeal Swabs
Authors:
- Annette AudigéAAAnnette Audigé
- Jürg BöniJBJürg Böni
- Peter W. SchreiberPSPeter W. Schreiber
- Thomas ScheierTSThomas Scheier
- Roberto BuonomanoRBRoberto Buonomano
- Alain RudigerARAlain Rudiger
- Dominique L. BraunDBDominique L. Braun
- Gerhard EichGEGerhard Eich
- Dagmar I. KellerDKDagmar I. Keller
- Barbara HasseBHBarbara Hasse
- Christoph BergerCBChristoph Berger
- Huldrych F. GünthardHGHuldrych F. Günthard
Open Access
- 10 August 2021
- journal article
- research article
- Published by MDPI AG in Microorganisms
- Vol. 9 (8) , 1700
Abstract
Early identification and isolation of SARS-CoV-2-infected individuals is central to contain the COVID-19 pandemic. Nasopharyngeal swabs (NPS) serve as a specimen for detection by RT-PCR and rapid antigen screening tests. Saliva has been confirmed as a reliable alternative specimen for RT-PCR and has been shown to be valuable for diagnosing children and in repetitive mass testing due to its non-invasive collection. Combining the advantages of saliva with those of antigen tests would be highly attractive to further increase test capacities. Here, we evaluated the performance of the Elecsys SARS-CoV-2 Antigen assay (Roche) in RT-PCR-positive paired NPS and saliva samples (N = 87) and unpaired NPS (N = 100) with confirmed SARS-CoV-2 infection (Roche cobas SARS-CoV-2 IVD test). We observed a high positive percent agreement (PPA) of the antigen assay with RT-PCR in NPS, reaching 87.2% across the entire cohort, whereas the overall PPA for saliva was insufficient (40.2%). At Ct values ≤ 28, PPA were 100% and 91.2% for NPS and saliva, respectively. At lower viral loads, the sensitivity loss of the antigen assay in saliva was striking. At Ct values ≤ 35, the PPA for NPS remained satisfactory (91.5%), whereas the PPA for saliva dropped to 46.6%. In conclusion, saliva cannot be recommended as a reliable alternative to NPS for testing with the Elecsys Anti-SARS-CoV-2 Antigen assay. As saliva is successfully used broadly in combination with RT-PCR testing, it is critical to create awareness that suitability for RT-PCR cannot be translated to implementation in antigen assays without thorough evaluation of each individual test system.Keywords
Funding Information
- Swiss Federal Office of Public Health (NA)
- University of Zurich Foundation (NA)
This publication has 56 references indexed in Scilit:
- Saliva sample as a non-invasive specimen for the diagnosis of coronavirus disease 2019: a cross-sectional studyClinical Microbiology and Infection, 2021
- Comparison of SARS-CoV-2 detection in nasopharyngeal swab and salivaJournal of Infection, 2020
- Saliva as an Alternate Specimen Source for Detection of SARS-CoV-2 in Symptomatic Patients Using Cepheid Xpert Xpress SARS-CoV-2Journal of Clinical Microbiology, 2020
- Saliva as a Noninvasive Specimen for Detection of SARS-CoV-2Journal of Clinical Microbiology, 2020
- Saliva is a reliable tool to detect SARS-CoV-2Journal of Infection, 2020
- Saliva—Friend and Foe in the COVID-19 OutbreakDiagnostics, 2020
- Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort studyThe Lancet Infectious Diseases, 2020
- Consistent Detection of 2019 Novel Coronavirus in SalivaClinical Infectious Diseases, 2020
- Evaluating the use of posterior oropharyngeal saliva in a point-of-care assay for the detection of SARS-CoV-2Emerging Microbes & Infections, 2020
- Clinical Significance of a High SARS-CoV-2 Viral Load in the SalivaJournal of Korean Medical Science, 2020