Technical note: Validation of the BHBCheck blood β-hydroxybutyrate meter as a diagnostic tool for hyperketonemia in dairy cows
- 1 February 2018
- journal article
- research article
- Published by American Dairy Science Association in Journal of Dairy Science
- Vol. 101 (2) , 1524-1529
Abstract
Accurate cow-side blood β-hydroxybutyrate (BHB) detection meters are valuable tools for rapid diagnosis of hyperketonemia. The main objective of this study was to compare the blood BHB measured in whole blood by the BHBCheck meter (PortaCheck, Moorestown, NJ) to a previously validated meter, Precision Xtra meter (Abbott Laboratories, Abbott Park, IL) and a colorimetric laboratory assay. Samples (n = 426) were collected from postpartum primiparous and multiparous Holstein cows (n = 79 cows) enrolled in 1 of 2 experiments (Exp) with different sampling schedules (Exp 1: n = 39 cows, 58 samples; Exp 2: n = 40 cows, 368 samples). In both Exp, whole-blood samples were collected from the coccygeal vessels after morning milking, before morning feeding. Blood samples were used immediately for BHB quantification via the BHBCheck meter and the Precision Xtra meter. Blood was also collected into evacuated tubes containing no additive (Exp 1) or potassium oxalate/sodium fluoride (Exp 2), which were centrifuged for serum or plasma separation and stored at -20°C for subsequent analysis. Laboratory quantification of BHB concentration was done by the BHB LiquiColor Assay (EKF Diagnostics-Stanbio, Boerne, TX; certified for serum and plasma). Data were analyzed by UNIVARIATE, CORR, FREQ, REG, and LOGISTIC procedures of SAS 9.4 (SAS Institute Inc., Cary, NC). Within this sample set, average parity was 3.3 lactations and DIM was 14 d. The proportion of samples classified as hyperketonemia (BHB ≥1.2 mmol/L) was 25, 28, and 31% as determined by the colorimetric assay, BHBCheck meter, and Precision Xtra meter, respectively. The correlation for BHBCheck meter BHB concentration compared with the colorimetric assay concentrations was r = 0.96, with a sensitivity of 91% and specificity of 93%. Correlation, sensitivity, and specificity of the Precision Xtra meter concentrations were 0.97, 98%, and 92%, respectively. Bland-Altman plots demonstrated minimal bias for both meters. Area under the receiver operator characteristic curve suggests adequate diagnostic accuracy of both meters. Overall, accuracy, sensitivity, and specificity of the BHBCheck meter was similar to the Precision Xtra meter and laboratory assay, indicating the BHBCheck meter is appropriate for use as a cow-side diagnostic test for hyperketonemia in dairy cows.Keywords
This publication has 16 references indexed in Scilit:
- Effects of a combination butaphosphan and cyanocobalamin product and insulin on ketosis resolution and milk productionJournal of Dairy Science, 2017
- Technical note: Comparison of 4 electronic handheld meters for diagnosing hyperketonemia in dairy cowsJournal of Dairy Science, 2016
- Ketosis Treatment in Lactating Dairy CattleVeterinary Clinics of North America: Food Animal Practice, 2013
- A field trial on the effect of propylene glycol on displaced abomasum, removal from herd, and reproduction in fresh cows diagnosed with subclinical ketosisJournal of Dairy Science, 2012
- Evaluation of an electronic cowside test to detect subclinical ketosis in dairy cowsJournal of Dairy Science, 2009
- Subclinical Ketosis in Lactating Dairy CattleVeterinary Clinics of North America: Food Animal Practice, 2000
- Ruminant Adaptation to Negative Energy BalanceVeterinary Clinics of North America: Food Animal Practice, 2000
- Biology of Dairy Cows During the Transition Period: the Final Frontier?Journal of Dairy Science, 1999
- Etiology of Lipid-Related Metabolic Disorders in Periparturient Dairy CowsJournal of Dairy Science, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986