Determination of nitrite ions in well water by biamperometric standard addition

Authors:

Abstract
A new modified biamperometric method of analysis is proposed which has the advantage of producing straight lines as plots. Nitrite ions in well water have been determined by biamperometry and standard addition of solutions containing nitrite or iodate ions. Two platinum foils are immersed in the solution and 100 mV potential is applied between the electrodes; the resulting current is measured. The standard nitrite or iodate ions react with excess of iodide ions, producing triiodide. The reversible redox pair I/I−is responsible for a change in current, allowing the biamperometric measurement. The concentration of triiodide formed depends on the nitrite concentration. Twenty ppb of nitrite ions have been determined with good accuracy and precision. EDTA was added to eliminate interference from Fe(III), Cu(II), and other metal ions.

This publication has 21 references indexed in Scilit: