Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes
Top Cited Papers
Open Access
- 31 May 2022
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 24 (6) , 373-380
Abstract
Background: We evaluated the accuracy and safety of a seventh generation (G7) Dexcom continuous glucose monitor (CGM) during 10.5 days of use in adults with diabetes. Methods: Adults with either type 1 or type 2 diabetes (on intensive insulin therapy or not) participated at 12 investigational sites in the United States. In-clinic visits were conducted on days 1 or 2, 4 or 7, and on the second half of day 10 or the first half of day 11 for frequent comparisons with comparator blood glucose measurements obtained with the YSI 2300 Stat Plus glucose analyzer. Participants wore sensors concurrently on the upper arm and abdomen. Accuracy evaluation included the proportion of CGM values within 15% of comparator glucose levels >100 mg/dL or within 15 mg/dL of comparator levels ≤100 mg/dL (%15/15), along with the %20/20 and %30/30 agreement rates. The mean absolute relative difference (MARD) between temporally matched CGM and comparator values was also calculated. Results: Data from 316 participants (619 sensors, 77,774 matched pairs) were analyzed. For arm- and abdomen-placed sensors, overall MARDs were 8.2% and 9.1%, respectively. Overall %15/15, %20/20, and %30/30 agreement rates were 89.6%, 95.3%, and 98.8% for arm-placed sensors and were 85.5%, 93.2%, and 98.1% for abdomen-placed sensors. Across days of wear, glucose concentration ranges, and rates of change, %20/20 agreement rates varied by no more than 9% from the overall %20/20. No serious adverse events were reported. Conclusions: The G7 CGM provides accurate glucose readings with single-digit MARD with arm or abdomen placement in adults with diabetes. Clinicaltrials.gov: NCT04794478Keywords
This publication has 17 references indexed in Scilit:
- State of Type 1 Diabetes Management and Outcomes from the T1D Exchange in 2016–2018Diabetes Technology & Therapeutics, 2019
- A head‐to‐head comparison of personal and professional continuous glucose monitoring systems in people with type 1 diabetes: Hypoglycaemia remains the weak spotDiabetes, Obesity and Metabolism, 2018
- Performance of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System in Pediatric Participants With Type 1 DiabetesJournal of Diabetes Science and Technology, 2018
- Performance of a Factory-Calibrated Real-Time Continuous Glucose Monitoring System Utilizing an Automated Sensor ApplicatorDiabetes Technology & Therapeutics, 2018
- Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with DiabetesDiabetes Technology & Therapeutics, 2018
- Effect of Continuous Glucose Monitoring on Glycemic Control in Adults With Type 1 Diabetes Using Insulin InjectionsJAMA, 2017
- Continuous Glucose Monitoring vs Conventional Therapy for Glycemic Control in Adults With Type 1 Diabetes Treated With Multiple Daily Insulin InjectionsJAMA, 2017
- Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose SwingsJournal of Diabetes Science and Technology, 2015
- Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose MonitoringDiabetes Technology & Therapeutics, 2015
- New Features and Performance of a Next-Generation SEVEN-Day Continuous Glucose Monitoring System with Short Lag TimeDiabetes Technology & Therapeutics, 2009