Effect of scanning position and orientation on the readout of Gafchromic film scanned with two different light-emitting flatbed scanners
Authors:
- 22 October 2026
- journal article
- research article
- Published by Springer Nature in Radiation and Environmental Biophysics
- p. 1-11
Abstract
The present study investigates the dose readout characteristics of Gafchromic EBT3 films over a dose range of 0.5–5 Gy using two flatbed scanners in-built with different light sources, Epson Expression 11000XL and 12000XL. Film readouts were acquired repeatedly at varying scan positions and film orientations to evaluate scanner-related non-uniformities. At 0.5 Gy, the maximum variation in optical density over the entire scan frame was 15.5%, 13%, and 29%, for the red, green, and blue channels, respectively, for the Epson 11000XL, compared with 9.5%, 7%, and 22.3%, for the Epson 12000XL. At 5 Gy, the corresponding maximum differences were 2.5%, 4.0%, and 4.1%, for the 11000XL, and 6.7%, 2.5%, and 10.1%, for the 12000XL. Changes in film orientation relative to the scan direction introduced additional variability. For the Epson 11000XL, the highest orientation-related differences at 0.5 Gy and 5 Gy were 15.3%, 2.1%, and 10.4%, and, 6.3%, 3.6%, and 1.3%, for the red, green, and blue channels, respectively. In the Epson 12000XL, the corresponding differences were higher at low dose, reaching 8.2%, 7.0%, and 26.8% at 0.5 Gy, and 5.5%, 8.5%, and 4.7% at 5 Gy. The results demonstrate that scanner response is non-uniform across the scan area and is influenced by film orientation, position, irradiation dose, and scanner light source, with effects more pronounced at lower doses. LED-based scanners exhibit marginally superior performance compared with fluorescent lamp-based systems. Consistent film orientation and position during scanning is therefore recommended. A standard operating flowchart for scanner calibration and quality assurance is proposed.Keywords
This publication has 33 references indexed in Scilit:
- Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG‐55Medical Physics, 2020
- Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beamsHeliyon, 2018
- Evaluation of a LED-based flatbed document scanner for radiochromic film dosimetry in transmission modePhysica Medica, 2018
- Radiochromic film dosimetry: Past, present, and futurePhysica Medica, 2011
- Scanning orientation and polarization effects for XRQA radiochromic filmPhysica Medica, 2010
- Evaluation of the magnitude of EBT Gafchromic film polarization effectsAustralasian Physics & Engineering Sciences in Medicine, 2009
- Scanning orientation effects on Gafchromic EBT film dosimetryAustralasian Physics & Engineering Sciences in Medicine, 2006
- Temperature, humidity and time. Combined effects on radiochromic film dosimetersRadiation Physics and Chemistry, 1996
- Temperature and relative humidity dependence of radiochromic film dosimeter response to gamma and electron radiationRadiation Physics and Chemistry, 1995
- Temperature and humidity effects on the response of radiochromic dye filmsRadiation Physics and Chemistry (1977), 1981