Rapid acquisition of helium‐3 and proton three‐dimensional image sets of the human lung in a single breath‐hold using compressed sensing

Authors:

Abstract
Purpose: To develop and validate a method for acquiring helium‐3 (3He) and proton (1H) three‐dimensional (3D) image sets of the human lung with isotropic spatial resolution within a 10‐s breath‐hold by using compressed sensing (CS) acceleration, and to assess the fidelity of undersampled images compared with fully sampled images.Methods: The undersampling scheme for CS acceleration was optimized and tested using3He ventilation data. Rapid 3D acquisition of both3He and1H data during one breath‐hold was then implemented, based on a balanced steady‐state free‐precession pulse sequence, by random undersampling of k‐space with reconstruction by means of minimizing the L1 norm and total variance. CS‐reconstruction fidelity was evaluated quantitatively by comparing fully sampled and retrospectively undersampled image sets.Results: Helium‐3 and1H 3D image sets of the lung with isotropic 3.9‐mm resolution were acquired during a single breath‐hold in 12 s and 8 s using acceleration factors of 2 and 3, respectively. Comparison of fully sampled and retrospectively undersampled3He and1H images yielded mean absolute errors 0.9.Conclusion: By randomly undersampling k‐space and using CS reconstruction, high‐quality3He and1H 3D image sets with isotropic 3.9‐mm resolution can be acquired within an 8‐s breath‐hold. Magn Reson Med 74:1110–1115, 2015.
Funding Information
  • Siemens Medical Solutions
  • the National Heart, Lung, and Blood Institute of the National Institutes of Health (R01 HL079077)

This publication has 15 references indexed in Scilit: