Active shielding to reduce low frequency disturbances in direct current near biomagnetic measurements
Authors:
- 1 May 1999
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (5) , 2465-2470
Abstract
Measurements of dc near biomagnetic fields are disturbed by low frequency noise that is not reduced sufficiently by most of the magnetically shielded rooms or gradiometers. For this reason an active shielding system has been developed at the Biomagnetic Center of the University of Jena. This work describes the principle of the active shielding system and demonstrates its properties concerning the attenuation of disturbing fields, frequency range, and some applications in biomedical measurements. We achieved a reduction of external low frequency magnetic fields by more than 50 dB and an attenuation of the field gradient by about 25 dB. This active shielding enables measurements of near dc biomagnetic fields in investigations of periinfarct depolarizations after ischemic stroke and spreading depression in migraine patients.Keywords
This publication has 6 references indexed in Scilit:
- Viability thresholds and the penumbra of focal ischemiaAnnals of Neurology, 1994
- Pathophysiology of the migraine auraBrain, 1994
- New results in active noise compensation for magnetically shielded roomsMeasurement Science and Technology, 1993
- Magnetoencephalography of focal cerebral ischemia in rats.Stroke, 1992
- Magnetic fields associated with spreading depression in anaesthetised rabbitsBrain Research, 1991
- SPREADING DEPRESSION OF ACTIVITY IN THE CEREBRAL CORTEXJournal of Neurophysiology, 1944