A Passively Suspended Tesla Pump Left Ventricular Assist Device
Authors:
- 1 November 2009
- journal article
- adult circulatory-support
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 55 (6) , 556-561
Abstract
The design and initial test results of a new passively suspended Tesla type left ventricular assist device blood pump are described. Computational fluid dynamics (CFD) analysis was used in the design of the pump. Overall size of the prototype device is 50 mm in diameter and 75 mm in length. The pump rotor has a density lower than that of blood and when spinning inside the stator in blood it creates a buoyant centering force that suspends the rotor in the radial direction. The axial magnetic force between the rotor and stator restrain the rotor in the axial direction. The pump is capable of pumping up to 10 L/min at a 70 mm Hg head rise at 8,000 revolutions per minute (RPM). The pump has demonstrated a normalized index of hemolysis level below 0.02 mg/dL for flows between 2 and 9.7 L/min. An inlet pressure sensor has also been incorporated into the inlet cannula wall and will be used for control purposes. One initial in vivo study showed an encouraging result. Further CFD modeling refinements are planned and endurance testing of the device.Keywords
This publication has 6 references indexed in Scilit:
- Analysis of Optimal Design Configurations for a Multiple Disk Centrifugal Blood PumpArtificial Organs, 1999
- Induction of Ventricular Collapse by an Axial Flow Blood PumpASAIO Journal, 1998
- The Need for Standardizing the Index of HemolysisArtificial Organs, 1994
- Evaluation of a Multiple Disk Centrifugal Pump as an Artificial VentricleArtificial Organs, 1993
- A multiple disk centrifugal pump as a blood flow deviceIEEE Transactions on Biomedical Engineering, 1990
- Spectrophotometry Studies on Blood Serum and Plasma: The Physical Determination of Hemoglobin and BilirubinAmerican Journal of Clinical Pathology, 1954