The use of Fourier-transform infrared spectroscopy for the quantitative determination of glucose concentration in whole blood
- 17 June 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (13) , 2023-2032
Abstract
Fourier-transform infrared transmission spectroscopy has been used for the determination of glucose concentration in whole blood samples from 28 patients. A 4-vector partial least-squares calibration model, using the spectral range 950-1200 cm(-1), yielded a standard-error-of-prediction of 0.59 mM for an independent test set. For blood samples from a single patient, we found that the glucose concentration was proportional to the difference between the values of the second derivative spectrum at 1082 cm(-1) and 1093 cm(-1). This indicates that spectroscopy at these two specific wavenumbers alone could be used to determine the glucose concentration in blood plasma samples from a single patient, with a prediction error of 0.95 mM.Keywords
This publication has 19 references indexed in Scilit:
- Analysis of the results of α-d-glucose Fourier transform infrared spectrum deconvolution: comparison with experimental and theoretical dataSpectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 1998
- Measurement of glucose and other analytes in undiluted human serum with near-infrared transmission spectroscopyAnalytica Chimica Acta, 1998
- Rapid and reliable spectral variable selection for statistical calibrations based on PLS-regression vector choicesAnalytical and Bioanalytical Chemistry, 1997
- Application of Molecular Spectroscopy in the Mid-Infrared Region to the Determination of Glucose and Cholesterol in Whole Blood and in Blood SerumApplied Spectroscopy, 1997
- FTIR and laser-Raman spectra of oligosaccharides in water: characterization of the glycosidic bondCarbohydrate Research, 1996
- Non-invasive glucose monitoringCurrent Opinion in Biotechnology, 1996
- Multivariate Determination of Glucose in Whole Blood Using Partial Least-Squares and Artificial Neural Networks Based on Mid-Infrared SpectroscopyApplied Spectroscopy, 1993
- Reagentless Near-Infrared Determination of Glucose in Whole Blood Using Multivariate CalibrationApplied Spectroscopy, 1992
- Multivariate determination of glucose in whole blood by attenuated total reflection infrared spectroscopyAnalytical Chemistry, 1989
- An Introduction to Multivariate Calibration and AnalysisAnalytical Chemistry, 1987