Performance of a Haptic Feedback Grasper in Laparoscopic Surgery: A Randomized Pilot Comparison With Conventional Graspers in a Porcine Model
- 4 June 2019
- journal article
- research article
- Published by SAGE Publications in Surgical Innovation
- Vol. 26 (5) , 573-580
Abstract
Background. Compared with open surgery, minimally invasive surgery is limited by reduced sensation of tissue properties. A laparoscopic grasper with integrated haptic feedback technology that improves the ability to sense tissue properties might provide a solution. The force reflecting operation instrument (FROI) is a new laparoscopic grasper, designed to provide information about the interaction forces between the instrument and tissue through resistance in the handle. This pilot study aimed to assess the functionality of the FROI compared with a conventional grasper in an in vivo setting. Methods. In this randomized trial, we used a standard laparoscopic surgical setup to perform laparoscopic surgery in pigs. In all, 11 surgeons performed colorectal, gynecological, or urological procedures, once with the FROI and once with a conventional grasper. Participants were asked to complete the NASA Task Load Index Rating Scale and rate 5 specific features for both graspers. To capture opinions on the overall functionality of the FROI, participants were asked to answer 8 open questions. Results. The surgeons reported that the use of the FROI significantly improved tissue consistency perception, arterial pulse detection, and force control compared with the conventional grasper. No significant differences were found in surgeons’ muscular strain or operative time. The most emphasized topics in the open questions were improved soft-tissue handling and importance for complex procedures. Conclusion. Through this first in vivo analysis of the functionality of the FROI, a multispecialty group of laparoscopic surgeons confirmed the added value of haptic feedback technology in a live surgical setting.Keywords
This publication has 36 references indexed in Scilit:
- Hand-Assisted Approach to Laparoscopic Myomectomy and HysterectomyJournal of Minimally Invasive Gynecology, 2013
- The role of tactile feedback in grip force during laparoscopic training tasksSurgical Endoscopy, 2012
- Principles of sensorimotor learningNature Reviews Neuroscience, 2011
- Development and Validation of a Surgical Workload Measure: The Surgery Task Load Index (SURG‐TLX)World Journal of Surgery, 2011
- Artificial tactile feedback can significantly improve tissue examination through remote palpationSurgical Endoscopy, 2010
- Ergonomics of disposable handles for minimally invasive surgerySurgical Endoscopy, 2009
- Conversion in laparoscopic surgery: does intraoperative complication influence outcome?Surgical Endoscopy, 2009
- The value of haptic feedback in conventional and robot-assisted minimal invasive surgery and virtual reality training: a current reviewSurgical Endoscopy, 2009
- The optimal mechanical efficiency of laparoscopic forcepsSurgical Endoscopy, 2004
- Slip and damage properties of jaws of laparoscopic graspersSurgical Endoscopy, 2004