Background: Detection of IgG subclasses (IgGSc) is vital for the diagnosis and management of disease, especially IgG4‐related diseases (IgG4‐RD). This study aimed to evaluate the performances of the chemiluminescent immunoassay (CLIA) for detecting IgGSc and diagnosing IgG4‐RD by IgGSc.Methods: A total of 40 individuals with IgG4‐RD, 40 with primary Sjogren's syndrome (pSS), and 40 healthy controls (HCs) were enrolled. Serum samples were collected for the simultaneous detection of IgG1, IgG2, IgG3, and IgG4 by the Siemens immunonephelometric assay and the CLIA. The correlation analysis was performed, and diagnostic value was analyzed by the receiver operating characteristic (ROC) curve.Results: Patients with IgG4‐RD had higher IgG4 (p < 0.001) and lower IgG1 (p < 0.001) than those with pSS, and HC. The results by the Siemens immunonephelometric assay and the CLIA showed a strong correlation in detecting IgG1, IgG2, IgG3, and IgG4 (r = 0.937, r = 0.847, r = 0.871, r = 0.990, all p < 0.001, respectively). The sum of IgG1, IgG2, IgG3, and IgG4 using two assays strongly correlated with total IgG by the IMMAGE 800 (r = 0.866, r = 0.811, both p < 0.001, respectively). For discriminating IgG4‐RD from pSS and HC, no significant differences were observed in CLIA IgG4 and Siemens immunonephelometric assay IgG4 (z = 0.138, p = 0.891), which provided the area under the curves (AUCs) of 0.951 (p < 0.001) and 0.950 (p < 0.001), respectively. The AUCs of CLIA IgG1 and Siemens immunonephelometric assay IgG1 in distinguishing pSS from IgG4‐RD and HC were 0.761 (p < 0.001) and 0.765 (p < 0.001), respectively, with no significant differences (z = 0.228, p = 0.820).Conclusions: The CLIA and the Siemens immunonephelometric assay appeared to have good consistency with comparable diagnostic value in detecting IgGSc, especially IgG4, and IgG1 that can accurately identify IgG4‐RD or pSS in clinical practice.Keywords
This publication has 29 references indexed in Scilit:
- Clinical utility of serum IgG4 measurementClinica Chimica Acta, 2020
- Causes of hypereosinophilia in 100 consecutive patientsEuropean Journal of Haematology, 2020
- Immunoglobulin G and immunoglobulin G subclass concentrations differ according to sex and raceAnnals of Allergy, Asthma & Immunology, 2020
- A light-initiated chemiluminescent assay for rapid quantitation of allergen-specific IgG4 in clinical samplesClinica Chimica Acta, 2019
- IgG subclasses quantitation: Analytical performance of The Binding Site SPA PLUS ® human assay and comparison with Siemens BNII® assayClinical Biochemistry, 2018
- Utility of Serum IgG4 Levels in a Multiethnic PopulationThe American Journal of the Medical Sciences, 2018
- Similar but not consistent: Revisiting the pitfalls of measuring IgG subclasses with different assaysJournal of Clinical Laboratory Analysis, 2017
- Detection of Polyclonal Increases in Immunoglobulin G4 Subclass by Distinct Patterns on Capillary Serum Protein ElectrophoresisAmerican Journal of Clinical Pathology, 2016
- Elevated IgG4 Serum Levels Among Primary Sjögren's Syndrome Patients: Do They Unmask Underlying IgG4‐Related Disease?Arthritis Care & Research, 2013
- H CHAIN SUBGROUPS OF MYELOMA PROTEINS AND NORMAL 7S γ-GLOBULINJournal of Experimental Medicine, 1964