Compensation of signal loss due to cardiac motion in point‐resolved spectroscopy of the heart
Open Access
- 20 November 2013
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 72 (5) , 1201-1207
Abstract
Purpose: To study the influence of cardiac motion on point‐resolved single voxel spectroscopy (PRESS) of the human heart and to propose an improved sequence design to compensate for cardiac motion–related signal losses.Methods: Numerical simulations and in vivo measurements were performed to study motion effects in cardiac spectroscopy. Based on the findings, an improved PRESS sequence with reduced free induction decay spoiling areas was implemented. Myocardial triglyceride (TG) content as estimated with the improved PRESS sequence was compared with the default PRESS sequence commonly used for spectroscopy in nonmoving tissues that was provided by the manufacturer.Results: Numerical simulations and in vivo data show the importance of optimal trigger delays to avoid signal losses in cardiac spectroscopy. The improved PRESS sequence was found to yield a significantly reduced sensitivity to cardiac motion and trigger delay setting and provided a 1.7‐fold increase in signal‐to‐noise ratio on average compared with the default PRESS sequence. Intrasubject variation of myocardial TG content was reduced with the improved sequence compared with the default PRESS sequence (coefficient of variation: 0.11 ± 0.03 versus 0.23 ± 0.06).Conclusion: Improvement of the sequence design is essential to reduce the sensitivity of PRESS to cardiac motion. Magn Reson Med 72:1201–1207, 2014.Keywords
This publication has 22 references indexed in Scilit:
- Undersampled Cine 3D tagging for rapid assessment of cardiac motionJournal of Cardiovascular Magnetic Resonance, 2012
- Rapid quantification of myocardial lipid content in humans using single breath‐hold 1H MRS at 3 TeslaMagnetic Resonance in Medicine, 2011
- Longitudinally and circumferentially directed movements of the left ventricle studied by cardiovascular magnetic resonance phase contrast velocity mappingJournal of Cardiovascular Magnetic Resonance, 2010
- k‐space analysis of point‐resolved spectroscopy (PRESS) with regard to spurious echoes in in vivo1H MRSNMR in Biomedicine, 2008
- On restoring motion‐induced signal loss in single‐voxel magnetic resonance spectraMagnetic Resonance in Medicine, 2006
- T1, T2 relaxation and magnetization transfer in tissue at 3TMagnetic Resonance in Medicine, 2005
- Observation of cardiac lipids in humans by localized 1H magnetic resonance spectroscopic imagingMagnetic Resonance in Medicine, 1994
- Comparison of localized proton NMR signals of skeletal muscle and fat tissue in vivo: Two lipid compartments in muscle tissueMagnetic Resonance in Medicine, 1993
- Spatial Localization in NMR Spectroscopy in VivoAnnals of the New York Academy of Sciences, 1987
- 1H NMR chemical shift selective (CHESS) imagingPhysics in Medicine & Biology, 1985