Multicenter performance evaluation of the Abbott Alinity hq hematology analyzer
- 7 June 2019
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 57 (12) , 1988-1998
Abstract
Background: Alinity hq (Abbott) is a new high-throughput hematology analyzer that exclusively employs optical principles for detecting and enumerating blood cells. It reports 29 parameters, including a six-part white blood cell (WBC) differential. The aim of this multicenter study was to evaluate the analytical and clinical performance of the Alinity hq. Methods: Complete blood count (CBC) results and morphological flagging were compared to that of CELL-DYN Sapphire (Abbott) and 2 × 200-cell manual differential results, on 1473 whole-blood samples from a well-defined patient population from three different clinical laboratories in the Netherlands. In addition, within-run and within-laboratory precision, linearity, limit of quantitation, carryover and sample stability were assessed. External quality assessment samples were also evaluated. Results: Data analysis demonstrated strong concordance of Alinity hq results with those of CELL-DYN Sapphire for all CBC parameters, except for basophil granulocytes. Alinity hq WBC differential showed high level of agreement with manual differential results and exhibited a better agreement with manual basophil results than CELL-DYN Sapphire. The sensitivity of the Alinity hq Blast flag was 57.6%, equal to the 57.6% sensitivity of the CELL-DYN Sapphire’s Blast Alert. When considering samples with ≥5% blasts, the sensitivity of the Alinity hq Blast flag was 70.0%. Analytical performance of Alinity hq was shown to be consistent with state-of-the-art (SOTA) performance characteristics. Conclusions: Alinity hq CBC measurands demonstrated good overall agreement with results obtained with CELL-DYN Sapphire, as well as manual WBC differential. The analytical and clinical performance characteristics of Alinity hq make it well suited for clinical laboratories.Keywords
This publication has 19 references indexed in Scilit:
- ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte countingInternational Journal of Laboratory Hematology, 2014
- Performance and Abnormal Cell Flagging Comparisons of Three Automated Blood Cell CountersAmerican Journal of Clinical Pathology, 2013
- Comparison of Automated Differential Blood Cell Counts From Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in Normal and Pathologic SamplesAmerican Journal of Clinical Pathology, 2013
- Howell-Jolly body interference in reticulocyte countsClinical Chemistry and Laboratory Medicine (CCLM), 2013
- Is the Blood Basophil Count Sufficiently Precise, Accurate, and Specific?American Journal of Clinical Pathology, 2012
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluidsClinical Chemistry and Laboratory Medicine (CCLM), 2010
- European Multi-Center Evaluation of the Abbott Cell-Dyn Sapphire Hematology AnalyzerLaboratory Hematology, 2006
- The International Consensus Group for Hematology Review: Suggested Criteria for Action Following Automated CBC and WBC Differential AnalysisLaboratory Hematology, 2005
- A Rapid and Accurate Closed-Tube Immunoassay for Platelets on an Automated Hematology AnalyzerAmerican Journal of Clinical Pathology, 2000
- Evaluating the CELL-DYN® 3500 Haematology Analyser in an acute general hospitalClinical and Laboratory Haematology, 1996