Improved pulse transit time estimation by system identification analysis of proximal and distal arterial waveforms
Authors:
- Caroline A. RickardsCRCaroline A. Rickards
- Guanqun ZhangGZGuanqun Zhang
- Ramakrishna MukkamalaRMRamakrishna Mukkamala
- Orestis VardoulisOVOrestis Vardoulis
- Theodore G. PapaioannouTPTheodore G. Papaioannou
- Nikolaos StergiopulosNSNikolaos Stergiopulos
- Mingwu GaoMGMingwu Gao
- N. Bari OlivierNON. Bari Olivier
- 1 October 2011
- journal article
- research article
- Published by American Physiological Society in American Journal of Physiology-Heart and Circulatory Physiology
- Vol. 301 (4) , H1389-H1395
Abstract
We investigated the system identification approach for potentially improved estimation of pulse transit time (PTT), a popular arterial stiffness marker. In this approach, proximal and distal arterial waveforms are measured and respectively regarded as the input and output of a system. Next, the system impulse response is identified from all samples of the measured input and output. Finally, the time delay of the impulse response is detected as the PTT estimate. Unlike conventional foot-to-foot detection techniques, this approach is designed to provide an artifact robust estimate of the true PTT in the absence of wave reflection. The approach is also applicable to arbitrary types of arterial waveforms. We specifically applied a parametric system identification technique to noninvasive impedance cardiography (ICG) and peripheral arterial blood pressure waveforms from 15 humans subjected to lower-body negative pressure. We assessed the technique through the correlation coefficient ( r) between its 1/PTT estimates and measured diastolic pressure (DP) per subject and the root mean squared error (RMSE) of the DP predicted from these estimates and measured DP. The technique achieved average r and RMSE values of 0.81 ± 0.16 and 4.3 ± 1.3 mmHg. For comparison, the corresponding values were 0.59 ± 0.37 ( P < 0.05) and 5.9 ± 2.5 ( P < 0.01) mmHg for the conventional technique applied to the same waveforms and 0.28 ± 0.40 ( P < 0.001) and 7.2 ± 1.8 ( P < 0.001) mmHg for the conventional technique with the ECG waveform substituted for the ICG waveform. These results demonstrate, perhaps for the first time, that the system identification approach can indeed improve PTT estimation.Keywords
This publication has 18 references indexed in Scilit:
- Monitoring non-invasive cardiac output and stroke volume during experimental human hypovolaemia and resuscitationBritish Journal of Anaesthesia, 2011
- Parametric estimation of pulse arrival time: a robust approach to pulse wave velocityPhysiological Measurement, 2009
- Physiological and Medical Monitoring for En Route Care of Combat CasualtiesJournal of Trauma: Injury, Infection & Critical Care, 2008
- Determination of Aortic Pulse Wave Velocity From Waveform Decomposition of the Central Aortic Pressure PulseHypertension, 2008
- 2007 Guidelines for the Management of Arterial HypertensionJournal Of Hypertension, 2007
- Aortic Pulse Wave Velocity as a Marker of Cardiovascular Risk in Hypertensive PatientsHypertension, 1999
- Determination of pulse wave velocities with computerized algorithmsAmerican Heart Journal, 1991
- Pulse Transit Time as an Indicator of Arterial Blood PressurePsychophysiology, 1981
- Dispersion and Attenuation of Small Artificial Pressure Waves in the Canine AortaCirculation Research, 1968
- A Method Using Induced Waves to Study Pressure Propagation in Human ArteriesCirculation Research, 1957