Reduction of Platelet Trapping in Membrane Oxygenators by Transmembraneous Application of Gaseous Nitric Oxide
Authors:
Open Access
- 1 May 1996
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 19 (5) , 291-293
Abstract
Bleeding during extracorporeal circulation (ECC) is often induced and/or aggravated by thrombocytopenia due to platelet-trapping in hollow fiber membrane oxygenators (HFMO). Nitric oxide (NO) has platelet anti-aggregating and dis-aggregating properties. In a paired system we tested whether gaseous NO, added to the gas compartment of one of two parallel running heparin-bonded HFMO attenuated platelet-trapping. Platelet consumption was markedly reduced in the NO-treated HFMO. These data strongly indicate that the application of gaseous NO could prove a new therapeutical approach to reduce bleeding during ECC, serving as a new way of preventing platelet loss, thus reducing the need for high systemic heparinization.Keywords
This publication has 9 references indexed in Scilit:
- Inhibition of Platelet Aggregation by Inhaled Nitric Oxide in Patients with Acute Respiratory Distress SyndromeAnesthesiology, 1995
- The platelet function defect of cardiopulmonary bypass [see comments]Blood, 1993
- Bleeding time prolongation and NO inhalationThe Lancet, 1993
- Regulation of Vascular Homeostasis by Nitric OxideThrombosis and Haemostasis, 1993
- Extracorporeal Membrane Oxygenation (ECMO) for Neonatal Respiratory Failure: Experience from the Extracorporeal Life Support Organization (ELSO)1992
- Endothelium-derived relaxing factor inhibits thrombin-induced platelet aggregation by inhibiting platelet phospholipase CBlood, 1992
- An L-arginine/nitric oxide pathway present in human platelets regulates aggregation.Proceedings of the National Academy of Sciences, 1990
- Physiopathology and Management of Coagulation during Long-term Extracorporeal Respiratory AssistanceThe International Journal of Artificial Organs, 1990
- Treatment of Acute Respiratory Failure by Extracorporeal Carbon Dioxide Elimination Performed with a Surface Heparinized Artificial LungAnesthesiology, 1987