Effect of transverse magnetic fields on a simulated in-line 6 MV linac
Authors:
- J St AubinJAJ St Aubin
- S SteciwSSS Steciw
- B G FalloneBFB G Fallone
- 3 August 2010
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (16) , 4861-4869
Abstract
The effects of a transverse magnetic field on an in-line side-coupled 6 MV linear accelerator are given. The results are directly applicable to a linac-MR system used for real-time image guided adaptive radiotherapy. Our previously designed end-to-end linac simulation incorporated the results from the axisymmetric 2D electron gun program EGN2w. However, since the magnetic fields being investigated are non-axisymmetric in nature for the work presented here, the electron gun simulation was performed using OPERA-3d/SCALA. The simulation results from OPERA-3d/SCALA showed excellent agreement with previous results. Upon the addition of external magnetic fields to our fully 3D linac simulation, it was found that a transverse magnetic field of 6 G resulted in a 45 +/- 1% beam loss, and by 14 G, no electrons were incident on the target. Transverse magnetic fields on the linac simulation produced a highly asymmetric focal spot at the target, which translated into a 13% profile asymmetry at 6 G. Upon translating the focal spot with respect to the target coordinates, profile symmetry was regained at the expense of a lateral shift in the dose profiles. It was found that all points in the penumbra failed a 1%/1 mm acceptance criterion for fields between 4 and 6 G. However, it was also found that the lateral profile shifts were corrected by adjusting the jaw positions asymmetrically.Keywords
This publication has 10 references indexed in Scilit:
- An integrated 6 MV linear accelerator model from electron gun to dose in a water tankMedical Physics, 2010
- The design of a simulated in‐line side‐coupled 6 MV linear accelerator waveguideMedical Physics, 2010
- Installation of the 1.5 T MRI accelerator next to clinical accelerators: impact of the fringe fieldPhysics in Medicine & Biology, 2009
- Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of conceptPhysics in Medicine & Biology, 2009
- First MR images obtained during megavoltage photon irradiation from a prototype integrated linac‐MR systemMedical Physics, 2009
- TU-C-M100F-01: Development of a Linac-MRI System for Real-Time ARTMedical Physics, 2007
- Validation of Monte Carlo generated phase-space descriptions of medical linear acceleratorsMedical Physics, 1999
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- A Monte Carlo model of photon beams used in radiation therapyMedical Physics, 1995
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995