Effect of shot number on the calculated apparent diffusion coefficient in phantoms and in human liver in diffusion-weighted echo-planar imaging

Authors:

Abstract
Purpose To show the signal intensity varies with shot number in diffusion‐weighted (DW) echo‐planar imaging (EPI) and affects apparent diffusion coefficient (ADC) calculation. Materials and Methods This prospective study was performed on 35 adult patients and 20 volunteers. Measurements were made on a 3T scanner using a breathhold DW spin‐echo EPI (SE EPI) sequence. Three protocols were used: A) eight consecutive shots at a fixed b‐value of 0 seconds/mm2 with TR = 1000 and 3000 msec; B) seven consecutive shots at b‐values = 0, 1000, 750, 500, 250, 100, 0 seconds/mm2 (in that order) with TR = 3500 msec; and C) seven consecutive shots (as in B) with TR = 1000, 1750, and 7000 msec. Results For protocol A, signal intensity decreased significantly from the first to second shot (Pb = 0 seconds/mm2 image was used. Using the first b = 0 seconds/mm2, the mean ADC was 15% higher than using the second b = 0 seconds/mm2 (Pb = 0 seconds/mm2 and the second b = 0 seconds/mm2 decreased as the TR increased. Conclusion The signal intensity can vary with shot number in SE EPI. For TR ≥ 3000 msec, steady‐state is attained after one shot. Using data acquired prior to steady‐state confounds the calculation of ADC values. J. Magn. Reson. Imaging 2009;30:547–553.