Detection of Explosives Vapors with a Portable Detector Based on Quartz-Crystal Microbalance
- 1 January 2007
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 19300395,p. 248-251
Abstract
This paper presents the development of a portable gas sensor device prototype able to detect explosives vapors as 2,4 dinitrotoluene (DNT). This detector is based on a single polymer coated quartz crystal microbalance and an appropriate data processing implemented on a microcontroller. Several polymeric materials have been chosen and their sensitivity performances have been evaluated to determinate the most suitable material for DNT detection. This device offers a great sensitivity for DNT exposures (150 Hz/ppm), but this one is also selective towards interfering solvents exposures (< 1 Hz/ppm), independently from temperature and moist conditions.Keywords
This publication has 5 references indexed in Scilit:
- Thermal 1/f noise from the theory of partitions: application to a quartz resonatorPhysica A: Statistical Mechanics and its Applications, 2003
- Study of the noise in adsorption–desorption phenomena using the Allan variance and a quartz microbalanceSensors and Actuators B: Chemical, 2000
- Cavitands as selective materials for QMB sensors for nitrobenzene and other aromatic vapoursSensors and Actuators B: Chemical, 1993
- The investigation of coating materials for the detection of nitrobenzene with coated quartz piezoelectric crystalsAnalytica Chimica Acta, 1986
- Detection of explosives with a coated piezoelectric quartz crystalAnalytical Chemistry, 1979