Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities
- 27 December 2018
- journal article
- research article
- Published by Taylor & Francis in Journal of Sports Sciences
- Vol. 37 (12) , 1411-1419
Abstract
This study compared heart rate (HR) measurements for the Fitbit Charge HR 2 (Fitbit) and the Apple Watch devices with HR measurements for electrocardiogram (ECG). Thirty young adults (15/15 females/males, age 23.5 ± 3.0 years) completed the Bruce Protocol. HR measurements were recorded from the ECG and both devices every minute. Average HR for each participant was calculated for very light, light, moderate, vigorous and very vigorous intensities based on ECG-measured HR. A concordance correlation coefficient (CCC) was calculated to examine the strength of the relationship between ECG measured HR and HR measured by each device. Relative error rates (RER) were also calculated to indicate the difference between each device and ECG. An equivalence test was conducted to examine the equivalence of HRs measured by devices and ECG. The Apple Watch showed lower RER (2.4-5.1%) compared with the Fitbit (3.9-13.5%) for all exercise intensities. For both devices, the strongest relationship with ECG-measured HR was found for very light PA with very high CCC (>.90) and equivalence. The strength of the relationship declined as exercise intensity increased for both devices. These findings indicate that the accuracy of real-time HR monitoring by the Apple Watch and Fitbit Charge HR2 is reduced as exercise intensity increases.Keywords
This publication has 20 references indexed in Scilit:
- TROIKA: A General Framework for Heart Rate Monitoring Using Wrist-Type Photoplethysmographic Signals During Intensive Physical ExerciseIEEE Transactions on Biomedical Engineering, 2014
- Comparison of FitBit® Ultra to ActiGraph™ GT1M for Assessment of Physical Activity in Young Adults During Treadmill WalkingThe Open Sports Medicine Journal, 2014
- Tachycardia-induced cardiomyopathyThe American Journal of Medicine, 2003
- Exercise Standards for Testing and TrainingCirculation, 2001
- Validation of the Polar Heart Rate Monitor for Assessing Heart Rate During Physical and Mental StressJournal of Psychophysiology, 2000
- Exercise Intensity and Longevity in MenJAMA, 1995
- Skin photoplethysmography — a reviewComputer Methods and Programs in Biomedicine, 1989
- Heart Rate and Exercise Intensity During Sports ActivitiesSports Medicine, 1988
- A comparison of the Two One-Sided Tests Procedure and the Power Approach for assessing the equivalence of average bioavailabilityJournal of Pharmacokinetics and Biopharmaceutics, 1987
- Maximal oxygen intake and nomographic assessment of functional aerobic impairment in cardiovascular diseaseAmerican Heart Journal, 1973