An investigation into the reduction in flow rate of intravenous fluid by antireflux valves
Authors:
- 29 June 2005
- journal article
- Published by Wiley in Anaesthesia
- Vol. 60 (8) , 797-800
Abstract
Summary: We have investigated the extent to which antireflux (non‐return) valves impede the maximum flow rate of crystalloid in intravenous infusion systems when administered either under gravity or with 300 mmHg added pressure. When used with a 16G cannula, a reduction in flow rate of 19–38% was consistently seen. With a 20G cannula, no significant reduction in flow was found. The valve tested that produced the least reduction in flow was the Westcott Sae‐flo ® valve.Keywords
This publication has 2 references indexed in Scilit:
- An evaluation of methods of increasing the flow rate of i.v. fluid administrationBritish Journal of Anaesthesia, 1995
- Flow through intravenous cannulaeAnaesthesia, 1987