An evaluation of the SPIFE® 3000 semi‐automated gel electrophoresis system for the identification of hemoglobin variants and comparison of relative electrophoretic mobilities with manual gel electrophoresis methods
Authors:
- 19 April 2010
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 32 (3) , 307-311
Abstract
Laboratory identification of hemoglobin (Hb) variants can involve multiple techniques. The use of semi-automated instruments that perform gel electrophoresis and staining, such as the SPIFE 3000 electrophoresis system, can greatly reduce the labor required for these commonly used techniques. We performed a comparison of the method involved in SPIFE 3000 system with those of manual gel electrophoresis. A total of 22 540 samples were analyzed using the SPIFE 3000, and compared with mobilities on cellulose acetate and citrate agar gels using standard manual methods. The results were compared using relative electrophoretic mobilities (REM). Of the 191 Hb variants identified, only 13 had REM that differed from manual electrophoresis when analyzed using the SPIFE 3000 system. One variant (Hb O-Indonesia) showed different mobility on both acid and alkaline gels, two (Hb E, Hb Sunshine Seth) on alkaline gel only, and 10 (Hbs N-Baltimore, N-Seattle, O-Arab, Shelby, Summer Hill, Tak, Hasharon, M-Iwate, Q-Iran, and Setif) on acid gels only. The SPIFE 3000 semi-automated electrophoresis system produces similar results when compared with those of standard manual electrophoresis methods.Keywords
This publication has 13 references indexed in Scilit:
- Comparison of Sebia Capillarys Capillary Electrophoresis With the Primus High-Pressure Liquid Chromatography in the Evaluation of HemoglobinopathiesAmerican Journal of Clinical Pathology, 2008
- Diagnostic Approach to HemoglobinopathiesHemoglobin, 2007
- Evaluation of a new Sebia kit for analysis of hemoglobin fractions and variants on the Capillarys® systemClinical Chemistry and Laboratory Medicine (CCLM), 2006
- Identification of Hemoglobin Variants by HPLCClinical Chemistry, 2005
- HPLC Retention Time as a Diagnostic Tool for Hemoglobin Variants and Hemoglobinopathies: A Study of 60000 Samples in a Clinical Diagnostic LaboratoryClinical Chemistry, 2004
- Capillary Electrophoresis of HemoglobinClinical Chemistry and Laboratory Medicine (CCLM), 2003
- HbVar: A relational database of human hemoglobin variants and thalassemia mutations at the globin gene serverHuman Mutation, 2002
- ABNORMAL HEMOGLOBINS: LABORATORY METHODSHemoglobin, 2001
- THE LABORATORY DIAGNOSIS OF HAEMOGLOBINOPATHIESBritish Journal of Haematology, 1998
- Measuring Relative Electrophoretic Mobilities of Mutant Hemoglos and Globin ChainsHemoglobin, 1978